Connection to a Circle Anesthesia Ventilator System
IMPORTANT: This guide is provided as a convenience and for General Information only. Never
use this product without clearly and thoroughly understanding the most recent revision of the
INOmax DS, INOmax DS
for specific, updated information regarding warnings, cautions, checklists, diagrams, and/or
instructions contained in this guide.
See applicable warnings and cautions on page 1 of this document.
1
16
Note:
1. With a circle anesthesia breathing circuit, the
INOmax DS / INOmax DS
system will perform as specified in the technical
specifications with fresh gas flow rates equal to or
greater than the patient minute volume.
2. The breathing circuit between the sample tee and
the patient Y should be between 6 and 12 inches
(150 -300mm) long: greater than 6 inches to
minimize the sampling of mixed inspired/expired
concentrations and less than 12 inches to help
ensure correct NO
3. For OR ventilation systems with the inspiratory
flow measurements at the inspiratory port of the
absorber, place the Injector Module upstream of
the inspiratory flow sensor.
Mallinckrodt Manufacturing LLC, 6603 Femrite Drive, Madison, WI 53718-6801 USA
For assistance contact Technical Support 877-566-9466.
and INOvent Operation Manuals. The Operation Manual is the source
IR
2
3
15
14
11b
/ INOvent delivery
IR
measurement.
2
4
INOmax DS
INOmax DS
IR
INOvent
13
12
11 11a
1.
2.
3.
4.
5.
6.
7.
8.
9.
10. Absorber inspiratory port
11. Injector Module
12. Injector module cable
13. Injector module tubing
14. Inspiratory hose
15. 22M / 15F X 22M / 15F Adapter
16. Sample Tee
5
6
10
9
8
Patient wye
Gas sample line
Sample line inlet
INOmax DS / INOmax DS
INOvent delivery system
Bellows assembly
Ventilator drive gas line
Ventilator (rear view)
Absorber
Absorber expiratory port
A. input
B. output
7
/
IR
Part No. 20731 Rev-01