7.4. Other Symptoms

Other common wiring problems can also cause the
gas detector to malfunction. Check below for
additional conditions that will cause gas detector
issues.
Symptom
Green
Check power supply. Check Wiring.
power
LED off
Alarms in
If you experience alarms in the absence of a
leak, try setting an alarm delay.
the
Perform a bump test to ensure proper
absence of
operation.
a leak
Table 7-4 Other Symptoms and Possible Causes