17 Chromatographic Checkout
Table 44 NPD Checkout Conditions (continued)
Oven
Initial temp
Initial time
Rate 1
Final temp
Final time
ALS settings (if installed)
Sample washes
Sample pumps
Sample wash volume
Injection volume
Syringe size
Solvent A pre washes
Solvent A post washes
Solvent A wash volume
Solvent B pre washes
Solvent B post washes
Solvent B wash volume
Injection mode (7693A)
Airgap Volume (7693A)
Viscosity delay
Inject Dispense Speed (7693A)
PreInjection dwell
PostInjection dwell
Manual injection
Injection volume
Data system
Data rate
7 If using a data system, prepare the data system to perform one run using the loaded
8 Start the run.
250
checkout method. Make sure that the data system will output a chromatogram.
If performing an injection using an autosampler, start the run using the data system, or
creating a one sample sequence and pressing
If performing a manual injection (with or without a data system):
a Select
to prepare the inlet for splitless injection.
60 °C
0 min
20 °C/min
200 °C
3 min
2
6
8 (maximum)
1 µL
10 µL
2
2
8
0
0
0
Normal
0.20
0
6000
0
0
1 µL
5 Hz
on the touchscreen.
Operation Manual

Hide quick links:

Advertisement