Ramsey Medical petMAP+ II Operator's Manual page 9

Ecg
Hide thumbs Also See for petMAP+ II:
Table of Contents

Advertisement

 
    Do not immerse the ECG Module in water or any liquid.  If the 
Module is accidentally wetted, it should be thoroughly dried before use.  If 
the ECG Module has been dropped into water or other fluid, it should not 
be used and should be returned to the manufacturer for service. 
    If the ECG Module has been dropped or damaged in any way, it 
should be checked by qualified service personnel to ensure proper 
operation prior to use. 
 
    Use of accessories other than those specified by Ramsey Medical 
may result in danger to the patient, malfunction of the monitoring unit, 
and additional increased electro‐magnetic (EM) emissions or decreased EM 
immunity of the device. 
 
    Follow local governing ordinances and recycling instructions 
regarding disposal and recycling of device components and packaging. 
 
 
The petMAP+II is not intended for use on patients being imaged 
with an MRI device since the petMAP device contains magnetically active 
materials and could result in injury if used too close to an MRI device, even 
if the MRI is not operating since the MRI magnet is still active even if not in 
use.   
 
Do not gas sterilize or autoclave the device. 
 
 
 