Ramsey Medical petMAP+ II Operator's Manual page 8

Ecg
Hide thumbs Also See for petMAP+ II:
Table of Contents

Advertisement

   
Use of an Electrosurgical Unit (ESU) will often result in interference 
on the ECG trace and potentially erroneous heart rates.  Never use the ECG 
for determining heart rate until the ECG waveform has been stable for 10 
seconds or more after ESU use. The BP and SpO2 are typically not 
adversely affected by ESU usage, but under certain circumstances the ESU 
could adversely affect them.  Always assure that the ESU ground plate is 
well placed on the patient and placed distant from the ECG electrodes.  
 
   
Inaccurate ECG heart rate readings may result when an 
electrosurgical unit (ESU) is used while monitoring with the petMAP+II. If 
this is suspected, discontinue use of the petMAP+II while the ESU is in use.  
When ESU is used, patient leads and wires should be away from the 
surgical operation site and other devices. This reduces the burning risk due 
to a poor connection of the ESU neutral electrode. 
 
 
The scientific literature describes interference from a number of 
electrical devices that can produce artifacts that look like normal or 
abnormal ECG waveforms but are not actual ECG waveforms, and these 
potential ECG artifact generators include IV pumps, line isolation monitors, 
electrosurgical units, pacemakers and other stimulators, fluid warmers, 
bronchoscopes, etc. In such instances, these devices are usually 
malfunctioning in some way when producing ECG artifacts, but not always. 
 
    In rare patient arrhythmias there may be 1 or more peaked 
complexes with each actual beat that resemble the actual, true QRS 
complex.  In such circumstances the unit can mistake the additional peaked 
complexes for a QRS and hence the heart rate displayed will be higher than 
the true heart rate. On the display where each QRS is detected there is a 
small indicator mark (yellow or red) below the detected QRS waveform to 
indicate it has been detected as a QRS.  If in doubt about the ECG heart 
rate, confirm that only true QRS complexes are being identified as such by 
looking at the small marks.  Even in the presence of such rare arrhythmias, 
the BP and the SpO2 heart rates should read properly. 
 

Advertisement

Table of Contents
loading

Table of Contents