Alarms

 
The following alarms can be generated by the ECG feature and displayed 
on the petMAP+II: 
No ECG Detected
 
O2 Sensor? ECG?
 
 
 
 
 
 
 
 
 
Situation:
 The ECG signal or
waveform amplitude is
insufficient to be detected.
Probable causes and/or
suggestions:
 The patient has no cardiac
electrical activity or rhythm.
 The amplitude or strength is
inadequate. Adjust the
electrodes, and/or increase
the gain.
NOTE: The "No ECG detected"
alarm becomes "O2 Sensor? ECG?"
if at the time the no ECG detected
alarm is triggered, there is already a
SpO2 alarm active.
 
or
 
ECG Lead OFF?
or
ECG: Lead? Switches?
Situation:
 Either a lead is off or the
ECG module has incorrect
switch settings.
Probable causes and/or
suggestions:
 Check the ECG leads.
Reposition electrodes and
add gel to surface leads if
necessary.
 Check to make sure that
both switches are correct:
o Esophageal or surface
o Gain