W&H Roto Quick RQ-03 Instructions For Use Manual page 25

Quick couplings
Table of Contents

Advertisement

pproved sterilization procedures
Follow your country-specific directives, standards and guidelines.
> Steam sterilization class B (pre and post vacuum) with sterilizers in accordance with EN 13060.
Sterilization holding time a minimum of 3 minutes at 134 °C (273.2 °F).
or
> Steam sterilization class S with sterilizers in accordance with EN 13060.
The sterilizer manufacturer must give their express approval for the sterilization Roto Quick couplings.
Sterilization holding time a minimum of 3 minutes at 134 °C (273.2 °F).
25