Pc & Storage Interface - Hanna Instruments Edge Instruction Manual

Multiparameter benchtop meter
Table of Contents

Advertisement

Logged data on edge™ can be transferred from the meter to a USB flash drive
by using the log recall function. The minimum requirement for the drive is USB
1.1. Select the pH, EC or DO record you wish to export and follow the simple
steps below.
Connect USB flash drive to the USB host socket,
located on the top of the meter. Press the RCL key
and select the parameter to view. Press the CFM key.
Select Manual, Stability, or interval lots by using the
p q arrow keys. Press the LOG key (not CFM). The
"USB HOST" tag should come on.
"PLEASE WAIT" message appears followed by
"EXPORT". Press CFM to export the selected record
or lot. If CFM is not pressed in 10 seconds, the USB
host will become inactive.
The meter will display the percentage of export.
The export percentage should go to 100%. Remove
USB flashdrive.
The display will return to the selected file. Press the
RCL key twice to return to measurements.
Note: Do not remove USB flash drive during an active
export transfer.
PC & Storage
Interface
Setup/Installation
25

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents