Installation; Environment Required; Install Spectrophotometer; Overview - Onda UV-20 User Manual

Table of Contents

Advertisement

contact us and authorized contractual partners.

Installation

1. Environment Required

To ensure the best performance, the following conditions are required:
The best working temperature range is 16—35℃ and the humidity
is 45—80%.
Keep it as far as possible away from the strong magnetic or
electrical fields or any electrical device that may generate
high-frequency fields.
Set the unit up in an area that is free of dust, corrosive gases
and strong vibrations.
Remove any obstructions or materials that could hinder the flow
of air under and around the instrument.
The power requirement is 110±11V/60±1Hz or 220±22V/50±1Hz.
Use the appropriate power cord and plug into a grounded outlet.
If the local voltage is not stable, a voltage regulator is required.
Be away from direct sunlight.

2. Install Spectrophotometer

Placement
Place the instrument on the stable table carefully.
Install Printer (Printer is Optional Accessories)
Check to confirm instrument power switch is turned off, connect the
printer's data cable to the instrument's parallel port.
Link the Power Cord
Check to confirm instrument power switch is turned off, the power cord
plug into two separate power interface and power supply socket apparatus.

Overview

UV-20 Spectrophotometer is an electrical measure instrument which is
widely used in the laboratories.
Use Frequency:
Excessive Voltage(Current):
Pollution Class:
Intermittence
No
Class 1
2