Dar k Cur r ent
Reset W L
Ver si on
O K

Troubleshooting

Review the information in the table below to troubleshoot operating
problems.
Problem
Power on, no response
Measurement uncertainty
Dark Current Error when
self-check
System Calibrate Failed
Cancel
to choose "Version", press
Cancel
Cause
Power cord connection is
not reliable
Fuse burning
Warm up is not enough
Sample is not Stable
The
concentration
sample is too high
Power Supply Voltage Low
or not Stable
Lamp damage or lamp life
maturity
The
lid
compartment
during self-check
Something
Light path
Cal i br at i ng W L!
SELECT
(left) to view version
Sof t war e: V2. 4. 2
Har dwar e: V1. 2
Solution
Improve connectivity
Replace fuse
Warm up more time
Improve the sample
of
Diluted sample
Improve the Power Supply
Replace lamp
of
the
is
open
Close the lid, restart
block
the
Remove
again
15
it,
calibrate