Repair And Maintenance; Daily Maintain; Spare Parts Replacement - Onda UV-20 User Manual

Table of Contents

Advertisement

Power on, back light is
OK, but nothing display
on the screen or display
is not clear
Measurements inaccurate

Repair and Maintenance

1. Daily Maintain

Check the Compartment
After measurement, the cuvettes with sample solutions should be taken out
of the compartment in time. Or the volatilization of the solution would
make the mirror go moldy. Users must pay more attention to the corrosive
sample and liquid easy to volatilize. Any solution remains in the
compartment should be wiped off immediately.
Surface Clean
The cover of the instrument is with paint. Please use wet towel to wipe
off the drips on the surface immediately. Organic solution is forbidden
to be used to clean the cover. Please wipe off the dirt on the cover timely.
Clean the Cuvettes
After every test or after a solution change, the cuvettes should be cleaned
carefully, or the remains on the surface would cause measuring error.

2. Spare Parts Replacement

Replace the Fuse
Danger! Be sure to switch off the power and unplug the socket
before replacement!
Step 1.
Tools preparation
Prepare a 3×75 Flat Blade screwdriver.
Display Contrast problem
Cuvettes
contaminated
Samples
contaminated
Worse matching of the
cuvettes
Dark current error
16
Adjust
the
potentiometer
were
Clean cuvettes
were
Improve samples
Improve the matching of
the cuvettes
Resample dark current
contrast