Installation

1. Environment Required

To ensure the best performance, the following conditions are required:
The best working temperature range is 16—35℃ and the humidity is 45—80%.
Keep it as far as possible away from the strong magnetic or electrical fields or any
electrical device that may generate high-frequency fields.
Set the unit up in an area that is free of dust, corrosive gases and strong vibrations.
Remove any obstructions or materials that could hinder the flow of air under and
around the instrument.
The power requirement is 110±11V/60±1Hz or 220±22V/50±1Hz.
Use the appropriate power cord and plug into a grounded outlet.
If the local voltage is not stable, a voltage regulator is required.
Be away from direct sunlight.

2. Install Spectrophotometer

Placement
Place the instrument on the stable table carefully.
Install Printer (Printer is Optional Accessories)
Check to confirm instrument power switch is turned off, connect the printer's data cable to the
instrument's parallel port.
Link the Power Cord
Check to confirm instrument power switch is turned off, the power cord plug into two separate
power interface and power supply socket apparatus.

Overview

UV-20 Spectrophotometer is an electrical measure instrument which is widely used in the
laboratories.
Use Frequency:
Excessive Voltage(Current):
Pollution Class:

Symbols

The following chart is an illustrated glossary of the symbols that are used in this manual.
Intermittence
No
Class 1
2

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents