INSTALLATION REQUIREMENTS

A standard 110/120 AC Volt, 15/60-Hertz cycle, 3 Amp
electrical outlet is required for the power switch for the 1500
and 1300 Models; no electricity is required for Model 1200
or 1100. A unit can typically be installed in less than three
hours, including plumbing by the in-house biomedical or
maintenance department.
For 1500, 1300 & 1200 Models, a 3/4" water line with a
minimum flow rate of 12 gallons per minute at 30 PSI is
required for optimal performance. For Model 1100, a 1/2"
water line with a minimum flow rate of 12 gallons per minute
at 30 PSI is required for optimal performance. Smaller pipes
or slower flow rates will result in longer evacuation times.
Purchaser of the Aqua-Box is responsible for the installation of
the included RPZ (reduced pressure zone backflow preventer)
valve for the fresh water line. The Model 1100 does not come
with a back flow preventer.

DISINFECTANT SOLUTION

This feature has been added mainly to accommodate those
facilities that simply prefer to treat their fluid on their own
accord. This feature can be easily turned on or off. The ratio
of sodium hypochlorite (common household bleach) to waste
fluid is pre-set at 1,100 PPM (parts per million). The Model
A q u a - B o x O w n e r ' s M a n u a l
1100 does not have a disinfectant solution inlet.
POWER SWITCH/TIMER
Models 1500-00 & 1300-00
The Aqua-Box power button is set to stay on for 2 minutes.
This pre-set time can be easily changed by the facility as
needed. A "kill" switch (the red button on the right side of
the electrical box) will turn off the unit prior to the end of the
timed cycle if desired.
Model 1300-R
The Aqua-Box power button is set to stay on for 5 minutes
This pre-set time can be easily changed by the facility as
needed. A "kill" switch (the red button on the right side of
the electrical box) will turn off the unit prior to the end of the
timed cycle if desired.
Model 1200-00 & 1100-00
These models do not have a power switch/timer.