Liner Systems - DeRoyal Aqua-Box 1200-00 Owner's Manual

Table of Contents

Advertisement

F l u i d W a s t e E v a c u a t i o n & D i s p o s a l
8. When canister is empty, leave thumb valve open for a
few seconds, then close thumb valve (D), replace cap on
canister, then discard empty canister.
9. If evacuating multiple canisters at one time, leave the
thumb valve (D) open and move the evacuation tube to the
next canister. Repeat until the last canister has been emptied.
10. When the last canister has been evacuated, and with the
Aqua-Box still running, place the Evacuation Tube in either:
a.
Fresh water for 5 seconds to flush the Aqua-Box.
b.
Enzymatic solution for 2 seconds to flush,
deodorize, and prevent organic and mineral deposits
from accumulating within the Aqua-Box.
11. Close the thumb valve (D) on the suction hose.
12. Turn the unit "Off".
a.
For Models 1500 and 1300, press the "Kill" switch
(B) on the side of the control box.
b.
For Models 1200 and 1100, turn the valve on the
right side of the unit to the "Off" position.
13. Discard the old evacuation tube and replace with a new
tube.
14. Hang up suction hose (C) on suction hose hanger (G).

LINER SYSTEMS

When purchasing an Aqua-Box for a facility that utilizes
liners without pour spouts, such as those manufactured by
Hospira, the Aqua-Box must be ordered with a liner adapter.
Because the adapter attaches directly to the liner portal, it
eliminates the need for evacuation tubes.
NOTE: A liner adapter can only be used with flexible liners.
1. Remove liner from canister shell and uncap tubing. Use a
clamp on the hose.
2. Press fit end of suction hose to any fluid port on liner.
(Be careful not to attach suction hose to vacuum port on
liner because fluid will be prevented from being evacuated
due to shut off valve).
3. Turn the unit "On".
a.
For Models 1500 and 1300, press the "Start" button
(A). If the Aqua-Box shuts off while evacuating,
simply press "Start" button (A) again.
b.
For Models 1200 and 1000, turn the valve on the
right side of the unit to the "On" position.
4. Liner will collapse as fluid is evacuated.
5. Detach the liner from the suction hose and replace tube
to suction port and discard.
6. With the Aqua-Box running, aspirate some disinfectant
through adapter.
7. Turn the unit "Off".
a.
For Models 1500 and 1300, press the "Kill" switch
(B) on the side of the control box.
b.
For Models 1200 and 1100, turn the valve on the
right side of the unit to the "Off" position.
17

Advertisement

Table of Contents
loading

This manual is also suitable for:

Aqua-box 1300-00Aqua-box 1500-00Aqua-box 1300-r

Table of Contents