Direct Connection; Coaxial Connection; Wire Extension - Siemens Echomax XCT Series Operation Manual

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Interconnection

Direct Connection

Connect the transducer directly to the Siemens Milltronics transceiver via the two conductor
shielded cable.
When connecting to an EnviroRanger ERS 500, a MultiRanger 100/200, or a
Note:
HydroRanger 200, the white, black, and shield wires are all connected separately. DO NOT
tie the white and shield wires together.

Coaxial Connection

Connect the transducer to the Siemens Milltronics transceiver via a junction box and RG–62 A/
U coaxial cable. This setup is effective for combined runs up to 365 m (1200 ft).
When connecting to an EnviroRanger ERS 500, MultiRanger 100/200, and
Note:
HydroRanger 200, do NOT use coaxial cable; see diagram below for proper procedure.

2-Wire Extension

(for EnviroRanger ERS 500, MultiRanger 100/200, and HydroRanger 200 only)
7ML19985QM82
XPS/XCT Series Transducers – OPERATION MANUAL
blk
wht
blk
wht
junction box
wht
drain/shield
drain/shield
extend cable using RG-62 A/U coax
extend cable using 18 AWG
shielded/twisted pair
drain/shield
Page EN-9

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents