13.1 Cleaning/decontamination
Cleaning of the METPOINT
one-way wipe, and mild, commercially available cleaner/soap.
For decontamination, spray the cleaner on an unused cotton cloth or one-way wipe and wipe the component
comprehensively. Effectuate subsequent drying using a clean cloth or via air drying.
In addition, the local hygiene provisions need to be observed.
Danger!
Compressed air/reaction products!
Risk of serious injury or death through contact with quickly or suddenly escaping compressed air
and through toxic, flammable, or potentially explosive reaction products.
Measures:
•
Only carry out cleaning measures when the system is pressureless.
•
Purge the respective plant component or pressure transducer prior to starting the maintenance works.
•
Immediately clean the removed components from media residues.
Warning!
Damage possible!
A too high degree of humidity and hard and pointed objects cause damage to the
pressure transducer and to the integrated electronic components.
Measures
•
Never clean with a soaked cloth.
•
Do not use pointed or hard objects for cleaning.
®
METPOINT
SP11
®
PRM SP11 must be undertaken using a slightly damp (not wet) cotton cloth or
Maintenance and calibration