Technical Data - Beko METPOINT PRM SP11 Instructions For Installation And Operation Manual

Pressure transducer
Table of Contents

Advertisement

8

Technical data

General indications
Model designation
Measuring principle
Measured parameter
Output signal, type PRM SP21
Measuring range
Overload pressure limit
Burst pressure
1
Process medium
Temperature-compensated range
Reference conditions
Process connection (connection pin acc. to EM 837-1)
Weight
Service life
Protection class acc. to EN 60529
Nonlinearity after limit point setting
Max. error of measurement
1
Process medium
Compressed air, nitrogen, water, oil, and other fluids of group 2 according to Article 9, Clause2.2 of the
PED 97/23/EC. Group 2 comprises fluids which are not toxic, flammable, potentially explosive, and fire-
promoting. The process medium must be compatible with the materials 1.4404 and 1.4548.
2
The indicated protection classes according to EN 60529 only apply when connected, with a mating
connector with a corresponding protection class.
3
Nonlinearity after the limit point setting, according to EN 61298-2
As regards the limit point setting, the reference line passes through the initial and end point
of the measured characteristic curve.
4
Max. error of measurement according to EN 61298-2
including nonlinearity, hysteresis, non-repeatability, and error of measurement of the measuring range final
value. Calibrated at a vertical installation position with a downward process connection.
5
MBE = Measuring range final value.
®
METPOINT
SP11
2
3
4
PRM SP11
Stainless-steel thin film technology
Overpressure (gauge pressure)
4 ... 20 mA, analogue, 2-wire
0 ... 16 bar
32 bar
125 bar
Gases/fluids
0 ... 60 °C
EN 61298-1
G¼ B
105 g
10 million load alterations
IP 67
5
≤ ± 0.15% MBE
5
≤ ± 0.5% MBE
Technical data
9

Advertisement

Table of Contents
loading