B. Braun Perfusor Space Service Manual page 7

Hide thumbs Also See for Perfusor Space:
Table of Contents

Advertisement

6
Procedural Instructions on the TSC
6 - 4
2. Battery check
Use the battery maintenance mode to discharge the battery
to battery pre-alarm (the battery status indicator displays no
bar).
Perform 5 syringe changes with emptied syringe.
-
Replace the battery if a battery alarm occurs during the
syringe changes.
3. Switch on unit with the power supply connected and check
the following details:
-
Self-test
Display on the LC display
-
Indicator lamps
LEDs (yellow, green, blue) light up for a short moment
LED changes from green to red
Changed order for units from SW J on:
LEDs (yellow, red, blue) light up for a short moment
LED changes from red to green
-
Audible alarm
A deep and a high sound
-
Status display
Battery capacity, mains operation (service connector SP)
-
Lighting
Illumination of syringe compartment, LC display and but-
tons
4. Syringe fastening
a) Axial fastening holds syringe wing
b) Syringe holder blade fixes syringe piston
c) Both claws in drive head catch the syringe piston plate
5. Check syringe recognition.
a) Insert approved 2 ml / 3 ml syringe.
The syringe size is recognized.
b) Insert approved 50 ml / 3 ml syringe.
The syringe size is recognized.
6. Carry out infusion and bolus with any syringe and press all
buttons at least once.
Infusion and bolus are performed and all buttons trigger the
function desired.