Charging The Battery; Troubleshooting - TensCare Flexistim IF Instructions For Use Manual

Hide thumbs Also See for Flexistim IF:
Table of Contents

Advertisement

12. CHARGING THE BATTERY

The Flexistim IF is powered by a type BL-6F rechargeable
Li-ion battery.
The battery cannot be charged while it is in the unit. A separate
charging cradle and power adaptor are included in the kit.
The battery will only last about one hour's use at 30mA.
When the battery is running low, a low battery indicator
will show on the screen (battery symbol).
NB: Remove the battery from your Flexistim IF if the unit is unlikely to be used for a
long period.
When the battery is charged, the indicator light on the cradle will change from red to
green. For a replacement battery, contact Tenscare or your local distributor.
Use only the power adaptor and charging cradle supplied.
The Flexistim IF can also be operated directly from the power supply. Just plug it directly
into the socket in the side of the unit. This automatically disconnects the battery.
USE OF OTHER CHARGERS COULD BE HAZARDOUS AND WILL NEGATE
THE GUARANTEE
Warning
There is a risk of smoke, fire, or rupture if the battery is not used according to the following
guidelines:
• Do not disassemble the battery
• Do not short-circuit the battery
• Do not incinerate or heat the battery
• Do not use or leave battery near a fire, stove or heated place (more than 80°C)
• Do not immerse the battery in water or sea water, or get it wet
• Do not charge battery nearby the fire or in strong sunlight
• Only use the charger provided and observe charging instructions
Disposal
Always dispose of batteries responsibly according to local government guidelines.
flexistimIF instructions final.indd 26-27
23

13. TROUBLESHOOTING

If your Flexistim IF is not working properly please check the following:
Problem:
No display/won't turn on:
BATTERY:
Controls don't work
No impulse output from electrodes
A circuit is not being made.
Warning triangle flashing,
When intensity reaches 40mA, the warning triangle
cannot increase intensity.
flashes and intensity cannot be increased.
Check electrode pad position. If you are certain that
the pads are positioned safely, press the SKIP key to
override. The triangle will stop flashing and intensity
can be increased to 60mA.
If the above review has failed to resolve your problem, call TensCare or your local dealer
for advice.
i) Is it fitted?
ii) Is it charged?
i) If
is shown on display, press button to unlock
the keypad
ii) No
showing. Ensure battery is charged.
i) Have you applied both electrode pads (per
lead wire) to ensure a complete circuit?
ii) Are the lead wires properly connected at
both ends?
iii) Is the lead damaged? (Try using the other
lead - if this works, then the original lead is faulty)