Thermal disinfection procedures - Manual
Disassembly
Disassemble the mask according to the instructions in the User Guide.
Cleaning
Alconox
AirFit F20 cushion
1. Soak the component in 1% Alconox solution at 68-77°F (20-25°C) for
1 minute.
2. Whilst immersed in the solution, clean with a soft bristle brush for 1
minute. Pay particular attention to all crevices and cavities.
3. Rinse by shaking vigorously in 5 liters of drinking quality water. Repeat
one more time.
Frame:
1. Rinse the component under running drinking quality water at 68-86°F
(20-30°C) for 1 minute.
2. Soak the component in 1% Alconox solution at 101-104°F (38.5-40°C)
for 5 minutes. Ensure no air bubbles are present and all surfaces come
in contact with the solution.
3. Whilst immersed in the solution, shake vigorously for another 5
minutes while squeezing the fabric of the arms.
4. Clean with a soft bristle brush for another minute. Pay particular
attention to all crevices and interfaces.
5. Rinse the component by agitating in 5 liters of drinking quality water at
66-68°F (19-20°C) for 1 minute.
6. Rinse the component under drinking quality water at 66-68°F (19-20°C).
Headgear:
1. Make a solution of Alconox by diluting with drinking quality water
at 1%, ie, 10 g per liter, at 68-86°F (20-30°C), according to the
manufacturer's instructions.
2. Soak the component in the solution for 3 minutes.
3. Thoroughly clean the mask component with a soft bristle brush for
2 minutes while continuing to soak it in the solution. Pay particular
attention to all crevices and cavities.
4. Squeeze the component to remove excess solution.
5. Rinse the component by repeatedly squeezing it under running drinking
quality water at 68-86°F (20-30°C) for 30 seconds. Repeat two more
times with fresh water.
Inspection / Drying
1. Inspect and if required, repeat washing until visually clean.
2. Allow the mask components to air dry out of direct sunlight.
Disinfection
1. Using a hot water disinfection system, soak the disinfectable mask components using a temperature-time combination:
• AirFit F20 cushion/elbow/frame: 167°F (75°C) for 30 minutes; 176°F (80°C) for 10 minutes; 194°F (90°C) for 1 minute;
• Headgear:
2. On completion, remove the mask components from the hot water disinfection system.
3. Allow the mask components to air dry out of direct sunlight.
Inspection
Perform a visual inspection of each mask component. If any visible deterioration of a mask component is apparent (cracking, crazing, tears etc), the mask component should be discarded and replaced. Slight
discoloration of the silicone components may occur and is acceptable.
Reassembly
Reassemble the mask according to the instructions in the User Guide.
Packaging and storage
Store in a dry, dust-free environment away from direct sunlight. Storage temperature: -4°F to 140°F (-20°C to 60°C).
These instructions are not intended for use with the AirTouch F20 cushion.
1
2
Failure to clean the mask component as indicated may result in inadequate disinfection and sterilisation.
:
1
2
200°F (93°C)
for 10 minutes
Elbow (manual):
1. Rinse the component under running drinking quality water at 59-77°F
(15-25°C) for 1 minute.
2. Soak the component in 1.5-2% Alconox solution at 104-122°F (40-50°C)
for 5 minutes. Ensure no air bubbles are present and all surfaces come
in contact with the solution.
3. Whilst immersed in the solution, shake vigorously for another 3
minutes. Brush the component for another minute. Pay particular
attention to swivel and vents.
4. Rinse the component by vigorously agitating in 5 liters of drinking
quality water at 59-77°F (15-25°C) for 1 minute.
5. Rinse the component under running drinking quality water at 59-77°F
(15-25°C) for 1 minute.
6. Shake the component to remove excess water.
Elbow (ultrasonic bath):
1. Rinse and clean the component with a soft bristle brush under running
drinking quality water at 68-86°F (20-30°C) for 1 minute. Pay particular
attention to all crevices, cavities and rotate moveable parts.
2. Fill two-thirds of an ultrasonic cleaner with 1% Alconox solution at
68-86°F (20-30°C) and sonicate for 20 minutes. Ensure the component
is fully immersed in the solution. Use a basket/tray to prevent contact
between the component and the tank.
3. Rinse the component by agitating vigorously in 5 liters of drinking
quality water at 68-86°F (20-30°C) for 1 minute. Repeat one more time.
200°F (93°C)
Control III
AirFit F20 cushion
:
1
1. Make a solution of Control III by diluting with
drinking quality water at 1.5-2%, ie, 15-20 g
per liter, at 68-86°F (20-30°C), according to
manufacturer's instructions.
2. Soak the component in the solution for 10-15
minutes.
3. Whilst immersed in the solution, clean with a
soft bristle brush. Pay particular attention to all
crevices and cavities.
4. Rinse by shaking vigorously in 5 liters of drinking
quality water. Repeat one more time.
for 10 minutes