Understanding Control Test Results

Your control solution test results should fall within the control solution range. If the results are within the
range, the GE006 Blood Glucose Monitoring System is working correctly.
Control Solution Range
Level 2
88~120 mg/dL
Possible reasons your Control Solution results are out of the range :
- Your Rightest™ Control Solution GC550 is expired or was first opened more than 3 months ago.
- Your GE006 Blood Glucose Test Strip has expired.
- You left the cap of the Blood Glucose Test Strip vial or the control solution off for a period of time.
- You did not perform the test procedure correctly.
- The GE006 Blood Glucose Meter or GE006 Blood Glucose Test Strip have malfunctioned.
If Rightest™ Control Solution GC550 results are out of range, your GE006 Blood Glucose Monitoring System
may not be working properly. Repeat the Quality Control Test. If your control solution results are still out of
range, do not use GE006 Blood Glucose Meter to test your blood glucose. please contact Bionime
Customer Service at 886 0800 241 688.
42
Level 4
250~340 mg/dL
The
GE006
Blood Glucose Meter is able to automatically store a maximum of 500 test results with time
and date. If your meter has stored 500 results, the newest test result will replace the oldest one.
To recall your test memory, start the meter without a test strip inserted.
1. Press the Main button to switch from the Time Mode to Memory Mode. You will
see the "mem" icon in the upper left corner of the display. The latest test result will
appear with the number "1".
2. Use either the Left or Right button on the side to review all previous results with
date and time. Results display from the most recent (Sequence no. "1") to the
oldest (Sequence no. "500") in the lower right hand corner of the screen.
3. The Quality Control Test result can be recalled from the saved data. When "
appears under the data, the test was performed using the
Solution GC550
. The control solution test result will not be used when recalling
test average results.

Recalling Test Results

"
Rightest™ Control
43