Press "SET" to enter the menu. The probe parameters menu contains the pa-
rameters, DO and DG, which characterize the transfer function of the linearized
thermistor control probe. These parameters may be adjusted to improve the ac-
curacy of the bath. This procedure is explained in detail in Chapter 10.
The probe parameters are accessed by pressing "SET" after the name of the pa-
rameter is displayed. The value of the parameter may be changed using the
"UP" and "DOWN" buttons. After the desired value is reached press "SET" to
set the parameter to the new value. Pressing "EXIT" will cause the parameter to
be skipped ignoring any changes that may have been made.
DO
This probe parameter refers to the temperature at which the control probe out-
put would be 0. Normally this is set for -25.229.
DG
This probe parameter refers to the temperature span of the probe between 0 and
100% output. Normally this is set for 186.974.

Operating Parameters

The operating parameters menu is indicated by,
Operating parameters menu
PAr
Press "SET" to enter the menu. The operating parameters menu contains the
cutout reset mode setting.

Cutout Reset Mode

The cutout reset mode determines whether the cutout resets automatically when
the bath temperature drops to a safe value or must be manually reset by the
operator.
The parameter is indicated by,
Cutout reset mode parameter
CtorSt
Press "SET" to access the parameter setting. Normally the cutout is set for
manual mode.
Cutout set for manual reset
Cto=rst
To change to automatic reset mode press "UP" and then "SET".
Cto=auto
Cutout set for automatic reset