Quick Start

This chapter gives a brief summary of the steps required to set up and operate
the 7009 bath. This should be used as a general overview and reference and not
as a substitute for the remainder of the manual. Please read Section 8 carefully
before operating the bath.
4.1

Set Up

Set up of the bath requires careful unpacking and placement of the bath, filling
the bath with fluid, installing the probe and connecting power. Consult Chapter
4 for detailed instructions for proper installation of the bath. Be sure to place
the bath in a safe, clean and level location.
Fill the bath tank with an appropriate liquid. For operation at moderate bath
temperatures, clean distilled water works well. Carefully pour the fluid into the
bath tank through the large rectangular access hole above the tank avoiding
spilling any fluid. The fluid must not exceed a height of 1/2 inch below the bath
lid.
The control probe must be inserted through the lid into the bath and plugged
into the socket at the back of the bath.
4.2

Power

Plug the bath power cord into a mains outlet of the proper voltage, frequency,
and current capability. See Section 3.1, Specifications for power details. Set the
"HEATER" switch on the front panel to position "LOW" and turn the bath on
using the front panel "POWER" switch. The bath will turn on and begin to heat
or cool to reach the previously programmed temperature setpoint. The front
panel LED display will indicate the actual bath temperature.
4.3

Setting the Temperature

In the following discussion a solid box around the word SET, UP, EXIT or
DOWN indicates the panel button while the dotted box indicates the display
reading. Explanation of the button or display reading are to the right of each
button or display value.
To view or set the bath temperature set-point proceed as follows. The front
panel LED display normally shows the actual bath temperature.
24.68 C
When "SET" is pressed the display will show the set-point memory that is cur-
rently being used and its value. Eight set-point memories are available.
Bath temperature display
Quick Start
11