COLD BATHS
• Refrigerated baths require that the condensing coil be cleaned periodi-
cally. Accumulation of dust and dirt on the condenser will result in pre-
mature failure of the compressor.
• This bath has been equipped with a brownout and over voltage protection
device as a safety feature to protect the system components.
• Mode of Operation: This bath needs to be plugged into the line voltage
for at least 2 minutes before operation. This is only necessary for the first
time the bath is energized or when it is moved from one location to an-
other. Turning the bath ON or OFF does not trigger the delay.
• If a High/Low voltage condition exists for longer than 5 seconds, the bath
de-energizes. An amber indicator on the back panel lights when this con-
dition exists.
• Re-energization is automatic upon correction of the fault condition and af-
ter a delay cycle of about 2 minutes. If a fault condition exists upon appli-
cation of power, the bath will not energize.
• Under and Over Voltage Protection at 230 VAC
1.3

Fluke Authorized Service Centers

Please contact one of the following authorized Service Centers to coordinate
service on your Fluke product:
Fluke Calibration
799 E. Utah Valley Drive
American Fork, UT 84003-9775
USA
Phone: +1.801.763.1600
Telefax: +1.801.763.1010
E-mail: support@Flukescientific.com
Fluke Nederland B.V.
Customer Support Services
Science Park Eindhoven 5108
5692 EC Son
NETHERLANDS
Voltage Cutout: ±12.5% (203 - 257 VAC)
Voltage Cut In: ±7.5% (213 - 247 VAC)