data logging.
NOTE3:When the recording year and model number is changed, the new folder will be also created
accordingly.
NOTE4: When SD card records 1 points, "SD" icon will flash one time, once the "SD" icon flashes three
times, it means the meter is under low battery, the recording will be failed if the battery is low.
Manual data logging (MAX 199 points)
1. Set the sampling rate to "0" seconds (Refer to "Setting the data logger sampling rate").
2. In the manual mode, data is logged when press and hold ADJ button and screen shows recorded points
"00X" in the temp. block with icon "MEM" flash in few seconds.
e.g. Recorded 1 points, then bottom screen shows "001".
3. Long press CAL button to clear data (removed MANUAL.csv), screen shows "CLr".
NOTE: While screen shows "Err" by long pressing CAL, it stands for no data can be cleared or SD card
isn't being inserted.
NOTE: Once clear the data by long press CAL, there is no way to recover the data. If you want to keep
the previous data, rename the file "MANUAL.csv" in /CDH-SD11/ MANUAL.csv is required.
4. Data directory in SD card : /CDH-SD11/ MANUAL.csv
Note: When manual data records full (199 points), logging will continue, but with new data overwriting
old. If you want to keep the previous data, rename the file "MANUAL.csv" in /CDH-SD11/
MANUAL.csv is required.
SD Data Transfer to PC
a)
Remove the SD card from the meter.
b)
Insert SD card directly into a PC SD card slot or use a SD card reader.
c)
Open the saved documents (CSV.) in the folder from the PC to get the data.
d)
File name /Product number/ Sample rate/ Recording point/ Start recording time/ End recording time/
Recording date/time /Recording parameters will be shown in the CSV. file.
e)
Data show "0" stands for no measured value during recording period.
Automatic Data logging
 Manual Data logging
10