ALCO-SENSOR FST
DEPRESS THE
BUTTON
OFF
TO SELECT THE PASSIVE
MODE
Test an Open Container
Test a Subject Passively
Selecting the Passive Testing Mode by pressing the
Passive Test. This will be indicated by the display first indicating the instrument's
current temperature (for passive testing the operating range is from 0°C to 50° C)
PAS
and then flashing the
display. Once you see this flashing
Sensor FST is prepared to capture a Passive Sample.
THIS MANUAL WILL DESCRIBE THREE WAYS TO CAPTURE A PASSIVE
SAMPLE. FIRST WE WILL DESCRIBE HOW TO COLLECT A PASSIVE SAMPLE
OVER AN OPEN CONTAINER. SECOND. WE WILL DESCRIBE HOW THE
INSTRUMENT IS DESIGNED TO AUTOMATICALLY COLLECT A SAMPLE OF
AIR FROM A SUSPECT WHO IS BLOWING IN THE DIRECTION OF THE
INSTRUMENT; AND FINALLY WE WILL DESCRIBE HOW TO COLLECT A
PASSIVE AMBIENT AIR SAMPLE FROM THE VICINITY OF A SUSPECTED
ALCOHOL USER.
To test a beverage for the presence of alcohol, attach the
put the instrument in the Passive Mode with "
The drink sniffer adapter allows the inlet port of the Alco-Sensor FST to extend
beyond the body of the instrument and provide better access for sampling the
alcohol vapor that sits just above the liquid in a glass. Proper use of the 'drink
sniffer' adapter will increase the sensitivity of the Alco-Sensor FST when testing the
gas (or headspace) above the surface of a liquid.
The ASFST, with the drink sniffer attached, should be positioned within one or two
inches above the surface of the liquid.
SNIFFER ADAPTER OR THE UNIT IN THE LIQUID)
button and a sample will be captured for analysis. A "
"E31"
result likely indicates the presence of alcohol in the beverage.
(status message)
nEG
A "
" result indicates that the alcohol content in the sample was negligible.
The best method to passively test a subject with the Alco-Sensor FST is to
automatically capture a sample from a subject who blows at the Passive Sampling
Cup. To accomplish this, the instrument must have a Passive Sampling Cup
attached and be in the Passive Mode with "
have his/her lips approximately 2" from the Passive Sampling Cup and the
subject must blow at the cup – see picture at left). In this mode, when the
instrument detects a consistent flow of breath reaching the sample inlet port, an
automatic sample will be collected for analysis. A "
presence of alcohol in the collected sample. A "
alcohol content in the sample was negligible. Have the subject blow in the
direction of the sample cup for as long as possible. Once you hear the instrument
click the sample has been taken, it will be analyzed and a result reported.
button will initiate a
OFF
PAS
'mouthpiece drink sniffer'
PAS
" flashing.
(CAUTION: DO NOT IMMERSE THE DRINK
. After several seconds press the
POS
" or "E29"
PAS
" flashing. (Note: the subject must
POS
" result indicates the
nEG
" result indicates that the
Alco-Sensor FST Operator Manual, Page 18 of 33, 11/30/2017