rcL)
Test Recall (
After the test result has been calculated the instrument will display the result for several seconds after which the
OFF
instrument will power
. If the operator wants to review the result, while the instrument is
the
button and then simultaneously press the
OFF
optional functions that the instrument can perform. The first item on the list is
this function, pressing the
performed along with an intermittent displayed
Depending upon the version of the Alco-Sensor FST, the result may be displayed in two digits, three digits
or 0 (Zero), PAS (Pass), CAU (Caution), FAıl (Fail), and the unit of measure may or may not be displayed.
PAS)
Passive Sampling (
Two common uses for this mode are to sample and determine if alcohol exists in the ambient air around a subject
or in the headspace over an unknown liquid substance. If alcohol is detected, a positive indication is displayed.
When performing a passive test of a subject, for best results, attach the Passive Sample Cup as shown in the
diagram below. If you are sampling the headspace of unknown liquid, the Passive Sampling Cup can be used, but
is not required.
Note: Certain versions of the Alco-Sensor FST have a pass code or security feature on some menu
functions. If you access a menu function and the instrument does not operate according to the
instructions in this manual, contact your Supervisor. If your Supervisor does not have a pass code key
Intoximeters will need to be contacted.
Performing a Passive Test
DURING POWER UP
DEPRESS AND HOLD THE
POWER
BUTTON WHILE
OFF
AT THE SAME TIME
PRESSING THE
BUTTON.
ON
RELEASE THE BUTTONS
rcL
ONCE
IS DISPLAYED
DEPRESS THE POWER
BUTTON TO SCROLL TO
PAS
THE
OPTION
ON
button will prompt the instrument to alternately display the result from the last test
OFF
rcL
.
This will turn the unit on and display the first option from a menu listing. The first
rcL.
option will be
When the
button is pushed it will cycle to the next menu item. If you continue to
ON
ON
push the
button, the instrument will cycle through the whole list of menu items
ON
and will eventually scroll back to the
passive test, cycle the message to the
button. The display will show the first menu item off a list of
rcL
rcL
option. For the purpose of performing a
PAS
option.
Alco-Sensor FST Operator Manual, Page 17 of 33, 11/30/2017
CONDUCTING A SUBJECT TEST
OFF
, momentarily press