If these suggestions do not correct the problem, discontinue use of any
device and contact your local Aerogen sales representative.
Table 4. Aerogen USB Controller System Troubleshooting
If this happens:
The 30 Minute or 6
Hour indicator light is
green, but aerosol is not
visible.
The 30 Minute or 6 Hour
indicator does not light
when On/Off power
button is pressed.
The 30 Minute or 6 Hour
indicator lights amber.
Longer than expected
treatment time e.g. 6
mL of Normal Saline
(0.9%) should take no
longer than 30 minutes
to nebulize.
Medication is left in
the nebulizer after
nebulization cycle.
The 30 Minute and 6
Hour indicators flash
amber alternatively
twice.
30
It could mean:
No medication in nebulizer.
Aerogen Pro has not been
properly cleaned.
Nebulizer used beyond life
of product.
There is no power to the
system.
The Aerogen USB
Controller cable is
incorrectly connected.
Nebulizer used beyond life
of product.
Aerogen Pro has not been
properly cleaned.
Nebulizer used beyond life
of product.
Nebulizer was not turned
on or connected to power.
Aerogen Pro has not been
properly cleaned.
Aerosol was not generated.
It may be time to replace
the nebulizer.
Internal error condition.
Aerogen
Try this:
Refill medication.
Clean Aerogen Pro.
See Warranty and Life of
Product.
Verify the Aerogen USB
Controller is securely attached
to a functioning power source.
Verify green power indicator
light on AC/DC is on.
Verify the Aerogen USB
Controller cable is correctly
connected to the nebulizer.
See Warranty and Life of
Product.
Clean the Aerogen Pro.
See Warranty and Life of
Product.
Ensure that the nebulizer
is connected to power and
turned on.
Clean the Aerogen Pro.
Verify aerosol is visible.
See Warranty and Life of
Product.
See Warranty and Life of
Product.
®