Instruction Manual
8.2

RS485 Command Protocol Summary

COMMAND
BRIEF DESCRIPTION
READ ION GAUGE
Read the current displayed
PRESSURE
pressure of the ion gauge in
Torr
SET ADDR OFFSET
Set the communications
(RS485) address offset
(upper nibble)
TURN IG ON/OFF
Power up the filament and
start reading pressure.
TURN DEGAS
Start a degas cycle or turn
ON/OFF
off degas.
SET EMISSION
Choose either 4mA or 100
CURRENT
µA emission current.
SET FILAMENT
Choose Filament 1 or 2.
SET OVER
Set 100 µA overpressure
PRESSURE
turn-off point.
SETPOINT
SET TRIP POINT
Set the 'turns on below' (+)
I
FOR RELAY
pressure point for relay
and set the 'turns off above'
(-) pressure point for relay
READ IG ON/OFF
Find out if filament is
STATUS
powered up and gauge is
reading.
InstruTech, Inc.
COMMAND SYNTAX
#xxRD<CR>
(e.g., #01RD<CR>)
where xx = the two digit
address of the device (node)
#xxSAuu<CR>
(e.g., #01SA10<CR>)
(1)
where uu=00 or 10 or 20 or
30
#xxIGy<CR>
(e.g., #01IG1<CR>)
y=1=Turn ON
y=0 Turn OFF
#xxDGy<CR>
(e.g., #01DG1<CR>)
Y=1= Turn ON
y=0= Turn OFF
#xxSEy<CR>
(e.g., #01SE1<CR>)
y=1=4mA
y=0=100μA
#xxSFy<CR>
(e.g., #01SF1<CR>)
y=1 for F1
y=2 for F2
#xxSOy.yyE-yy<CR>
(e.g., #01SO4.00E-02<CR>)
#xxSLzy.yyEzyy<CR>
I
(e.g., #01SL+4.00E+02<CR>)
(e.g., #01SL-5.00E+02<CR>)
I
.
For the 'z' preceding the
'y.yy' value, a '+' = 'turns on
below' the setpoint pressure
and a '-' = 'turns off above'
the setpoint pressure for
RLY I
#xxIGS<CR>
(e.g., #01IGS<CR>)
IGM401 Hornet
RESPONSE
*xx_y.yyEzyy<CR>
where y.yy = mantissa, z = sign of
the exponent, i.e., +/- and pp =
the exponent.
(e.g., *01_1.53E-06<CR>)
When IG is off:
*01_9.90E+09<CR>
*xx_PROGM_OK<CR>
*xx_PROGM_OK<CR>
y=0=IG OFF (clears any IG errors)
*xx_PROGM_OK<CR>
*xx_PROGM_OK<CR>
*xx_PROGM_OK<CR>
*xx_PROGM_OK<CR>
*xx_PROGM_OK<CR>
If the 'turns off above' (-)
pressure setpoint is less than the
value of the 'turns on below' (+)
setpoint, an error message of ?01
SYNTX ER<CR> will result.
*xx_0_IG_OFF<CR>
*xx_1_IG_ON_<CR>
Page 32