Bakeout - Instrutech Hornet User Manual

Hot cathode ionization vacuum gauge
Table of Contents

Advertisement

Instruction Manual
4

Bakeout

If desired, a chamber bake may be performed for new systems or after routine maintenance. The IGM401
sensor can be baked out to 200
elastomer O-rings, the maximum bakeout temperature is limited to the maximum temperature rating of the
elastomer O-ring. Ensure the temperature of the sensor tube and the vacuum fitting to the sensor is at the
same or above the chamber temperature. The electronic module must be removed from the sensor if the bake-
out temperature is to exceed 70
1. Turn off power to the IGM401.
2. Disconnect the cable from the IGM401.
3. Use a 3/32 in. size Hex key to remove the #4-40 socket head cap screws (SHCS) as shown below.
4. Detach the metal enclosure and the electronics from the sensor. Gently pull the electronics enclosure away
from the sensor using a gentle rocking motion.
5. The black plastic cap attached to the sensor and the end plate does not have to be removed for bakeout.
6. Perform bake out with the electronics removed.
7. Reattach the electronics enclosure. Reinstall the 4-40 SHCS (4 ea.)
InstruTech, Inc.
o
C as long as the sensor fitting uses metal seals. For sensor fittings using
o
C. To bake out the sensor use the following procedure:
IGM401 Hornet
Page 14