5.7

Factory-Set Default Parameters

The following is a summary of all factory-set default
SETUP DISP
- SET CONTRAST [Factory default = 10]
- MODE [Factory default = NORMAL]
- FLIP SCREEN [Factory default = NORMAL]
SETUP UNIT
- UNITS [Factory default = TORR]
- DEFAULTS [Factory default = MENU TO EXIT]
- GAS TYPE [Factory default = N2]
- CLR IG ERROR [Factory default = OFF]
- OVER PRESS 100 [Factory default = 5.00E-02 TORR]
- IG CNTL [Factory default = DIGI/RS485, (Factory default is DIGI or Digital Input, see requirement for switching from
DIGI to RS485 mode at
shown here in this section 5.7 will be modified; See
Menu choices when IG CNTL is set for FRONT PANEL.
SETUP IG
- DEGAS TIME [Factory default = 2]
- SENSITIVITY [Factory default = Actual Sensor sensitivity (nominal = 8-13)]
- FILAMENT NUM [Factory default = FILAMENT 1]
I
- RLY
LO TRP [Factory default = 1.00E-06 TORR]
I
- RLY
HI TRP [Factory default = 5.00E-06 TORR]
I
- RELAY
TEST [Factory default = OFF]
- FIL USAGE [Factory default = Displays to-date filament usage]
- AUTO IE EN [Factory default = DISABLE]
- AUTO IE TRIP [Factory default = DISABLE]
SETUP COMMS
- BAUD RATE [Factory default = 19200, 8, N, 1]
- ADDR [Factory default = 1]
- ADDR OFFSET [Factory default = 0]
SERVICE MENU
- INFO [Factory default = FIRMWARE VERSION]
- OP TIME [Factory default = Displays to-date operating time]
1
The above correspond to the Granville-Phillips® Series 354 Micro-Ion® module factory settings. If you are replacing a Micro-Ion® module you do not
need to change any of the parameters programmed in the factory. If you wish to make any changes refer to the next section for menu item explanations.
InstruTech, Inc.
DIGI/RS485
section, below)] If IG CNTL is set to FRONT PANEL, the Top Level Program Menu
1
values in the IGM401 display menu.
section 5.8.2., - FRONT
IGM401 Hornet
PANEL, for modified Top-Level Program
Page 19