BSH5000 Operation manual Chapter 1 Introduction Chapter 1 Introduction The dry bath incubator is controlled by sophisticated software, which can be widely used for the preservation and reaction of samples, the amplification of DNA, the pre-denaturation of the electrophoresis, blood serum coagulation and a variety of laboratory procedures.
BSH5000 Operation manual Chapter 3 Basic Instructions Chapter 3 Overview This chapter focuses on the introductions of the structure, keypads and key-functions of the instrument, as well as preparatory work prior to initial operation. 1. Structure overview Transparent cover Dual Block chamber...
BSH5000 Operation manual Chapter 3 Basic Instructions 2. Keypads ℃ ℃ Temp. Temp. Time Time External External Start Start Stop Stop 3. Display Mode The “Mode” button is used to toggle between the time and temperature setting. Down key: Used in both time and temperature mode to decrease the setting.
BSH5000 Operation manual Chapter 4 Operation Guide Chapter 4 Operation 1. Temperature and time setting Press the On/Off switch to power on the instrument. The instrument performs a self test followed by an alarm to signal that the product is ready for use.
BSH5000 Operation manual Chapter 4 Operation Guide Timer Operation (Advanced) The default setting of the instrument is designed so that the heater will continue to hold the set temperature even once time has expired and “oVEr” is showing on the time display.
Page 9
BSH5000 Operation manual Chapter 4 Operation Guide a) Power on the machine and ensure that the temperature on the display is less than 35°C. b) Inject olefin oil into one of the cone-shaped wells, and then put a thermometer/thermocouple into this well (make sure that the precision of the thermometer should be within 0.1℃...
Page 10
BSH5000 Operation manual Chapter 4 Operation Guide temperature to ensure the calibration accuracy. If the actual read out of thermometer is 39.6, modify the temperature to 39.6 by pressing or . Then press “Start/Stop” to confirm the input value.
Page 11
BSH5000 Operation manual Chapter 4 Operation Guide The upper displayed temperature (Ex. 32.0) is the current temperature measured by the External sensor. When the temperature reaches to 40° and is held as constant the decimal digit begins to flash. Notes: Please allow 20 minutes before reading the displayed actual value to ensure the calibration accuracy.
BSH5000 Operation manual Chapter 4 Operation Guide procedure will be cancelled and the changed value will have of no effect. 4.Temperature control by external sensor a) Insert the external temperature probe into a well of the drybath or directly into a sample tube.
BSH5000 Operation manual Chapter 5 Failure Analysis and Troubleshooting Chapter 5 Failure Analysis and Troubleshooting Problems and actions Common problem Possible cause Action(s) No power on the main Check power supply power plug plugged properly Faulty fuse Change fuse display...
BSH5000 Operation manual Annex Annex:Wiring Diagram ( ) for reference only POWER MAIN Switch BOARD BOARD J4 Heat J2 PT100 USB 1 US B J4 PT1000 HEAT PT1000 Block CONN BOARD — 12 —...