For a period of one (1) year following the date of delivery, BUFFALO FILTER warrants the Viro- Vac™ against any defects in material or workmanship. BUFFALO FILTER will repair or replace (at BUFFALO FILTER’S option) the same without charge, provided that routine maintenance as specified in this manual has been performed using replacement parts approved by BUFFALO FILTER.
OPERATING INSTRUCTIONS System Controls ViroSafe® Filter Instructions Set-up and Operation Specifications MAINTENANCE General Maintenance Information Cleaning Periodic Inspection Troubleshooting CUSTOMER SERVICE Equipment Return Ordering Information TERMS & WARRANTY Figure Title Page Control Panel Buffalo Filter ViroVac™ Operator’s Manual: Page 2...
The third stage filtration uses the highest grade virgin activated carbon, especially designed for Buffalo Filter for the removal and adsorption of odors and toxic gases produced by burning tis- sues. These harmful gases may constitute a health hazard to healthcare professionals who are subjected to prolonged exposure.
8. Upon request, BUFFALO FILTER will provide the following: Service and Repair Instructions, including Circuit Diagrams and Parts List 9. The fuses on the circuit board are to be serviced by an authorized BUFFALO FILTER techni- cian as follows: 100/120 VAC, 50/60 Hz use 10 Amp 250 Volt Fuse (Slo-Blo) 220/240 VAC, 50/60 Hz use 8 Amp 250 Volt Fuse (Slo-blo) Buffalo Filter ViroVac™...
• Disconnect the unit from the electrical outlet prior to inspecting system components. • The ViroVac™ system is only intended and suitable for the applications that are mentioned in the operating instructions. Buffalo Filter ViroVac™ Operator’s Manual: Page 5...
Page 7
ViroVac™ or shielding the loca- tion. • The use of ACCESSORIES other than those specified by BUFFALO FILTER, or sold by BUF FALO FILTER as replacement parts for internal components, may result in increased emissions or decreased immunity of the ViroVac™.
Page 8
• Do not block either the tubing or the filter. If either becomes occluded or significantly restrict- ed, the motor/blower may overheat and cause the unit to fail. • Using any other filter or accessory not supplied by BUFFALO FILTER may cause damage and/ or cause the system to be inoperable and may void the warranty.
Page 9
There are no user serviceable components in the ViroVac™ Smoke Evacuation System(s). Refer service to qualified service personnel. Use only with the power cord provided and always plug into a grounded outlet. Buffalo Filter ViroVac™ Operator’s Manual: Page 8...
A footswitch or a Remote Switch activator may be added to any system by simply plugging in a BUFFALO FILTER activation accessory into the appropriate jack on the front of the unit. When the footswitch is plugged in, the unit may be turned on or off by depressing the footswitch pedal once for each operation.
Page 12
If the system is overheated or if there is an electrical surge in the electrical system, fuses will break and the system will not operate. When the Service light illuminates, please contact BUFFALO FILTER Customer Service for system service instructions.
3. Clean the unit with appropriate germicide prior to re-use. Follow the indicated instructions for maintenance and installation of a new ViroSafe Filter. CAUTION: Using any other filter or accessory not supplied by BUFFALO FILTER may cause damage to the system and/or cause the system to be inoperable and may void the warranty.
Releasing either the CUT or COAG button on the Electrosurgical Pencil (If Remote Activa- tor is connected). 8. Replace the ViroSafe Filter when the Filter Life Remaining Scale FLASHES RED (0% Life Re- maining). Failure to change the filter will affect the performance of the system. Buffalo Filter ViroVac™ Operator’s Manual: Page 13...
Page 15
Figure 1 Unit Control Panel Buffalo Filter ViroVac™ Operator’s Manual: Page 14...
Section 3.0 General Maintenance Information This section contains information for ordinary upkeep of the BUFFALO FILTER ViroVac™ Smoke Evacuation System. While the system has been designed and manufactured to high industry standards, it is recommended that periodic inspection and performance testing be performed by a qualified Biomedical Technician to ensure continued safe and effective operation.
To reorder, obtain replacement parts or to return a unit for service, call Customer Service at: 800-343-2324 (716) 835-7000 or contact your authorized BUFFALO FILTER Distributor/Representative. BUFFALO FILTER ViroVac™ Smoke Evacuation System versions available: • 100/120 VAC 50/60 Hz • 220/240 VAC 50/60 Hz Available accessories: •...
(1) year of the date of delivery for P/N od. Buffalo Filter reserves the right to decide VV120 & VV220 and provided Buffalo Filter shipping method on prepaid orders. Care is...
Page 25
ENTIRE AGREEMENT: Purchaser by acceptance of Buffalo Filter’s of- fer does acknowledge and agree to the terms and conditions contained herein. All matters involving the validity, interpretation and ap- plication of this agreement shall be controlled by the laws of New York State. Using any...