ChoiceMMed MD300I Instruction Manual

ChoiceMMed MD300I Instruction Manual

Pulse oximeter

Advertisement

PULSE OXIMETER
INSTRUCTION
MANUAL
MD300I

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MD300I and is the answer not in the manual?

Questions and answers

Summary of Contents for ChoiceMMed MD300I

  • Page 1 PULSE OXIMETER INSTRUCTION MANUAL MD300I...
  • Page 2: Table Of Contents

    Copyright Content Our company owns all rights of this unpublished work and intends to maintain it as a 1 Introduction ....................1 confidential work. We may also seek to maintain this work as an unpublished copyright. This 1.1 Brief Introduction ........................1 publication is to be used solely for the purpose of reference or operation of our software 1.2 Intended Use ........................1 system.
  • Page 3: Introduction

    The device is designed of handheld structure and please be sure not to turn upside down when using it. 1.2 Intended Use The MD300I pulse oximeter is a portable non-invasive device intended for spot-checking of oxygen saturation of arterial hemoglobin (SpO ) and pulse rate of adult and pediatric patient at home and hospital (including clinical use in internist/surgery, Anesthesia, intensive care and etc).
  • Page 4 MD300I MD300I PULSE OXIMETER INSTRUCTION MANUAL • Note - Provides application tips or other useful information to ensure that you get the • This equipment is not intended for use during patient transport outside the healthcare most from your product.
  • Page 5: Electromagnetic Interference

    MD300I MD300I PULSE OXIMETER INSTRUCTION MANUAL adversely affect the accuracy of the SpO2 readings. this international standard are: CISPR11, GROP1, and CLASS B. • Obstructions or dirt on the sensor’s red light or detector may cause a sensor failure. 1.6 Explanation of Symbols Make sure there are no obstructions and the sensor is clean.
  • Page 6: Chapter 2 General Descriptions

    MD300I MD300I PULSE OXIMETER INSTRUCTION MANUAL ● OLED display screen with adjustable backlight. CHAPTER 2 General Descriptions ● Up to 10 patients’ IDs and 72-hour records can be saved. 2.1 Appearance ● Data transfer to PC by USB cable. ● Powered by two AA alkaline.
  • Page 7: Sensor Connection

    MD300I MD300I PULSE OXIMETER INSTRUCTION MANUAL Install the batteries with the correct polarity. Incorrect placement may cause damage to the CHAPTER 3 Take a Measurement bracket. Note: Definition of key press Please remove the batteries if the pulse oximeter will not be used for long periods of time.
  • Page 8: Warning

    MD300I MD300I PULSE OXIMETER INSTRUCTION MANUAL Low priority: indicates the technical alarm caused by the device itself. Alarms of the oximeter include technical and physiological alarms. All the three priorities are divided by built-in module and cannot be changed by user.
  • Page 9: Chapter 4 Settings

    MD300I MD300I PULSE OXIMETER INSTRUCTION MANUAL Holding the function key (left key) for more than one second, the brightness level headed with CHAPTER 4 Settings “Br” will be shown on the top right of the screen. You can adjust the brightness by degrees by pressing the setting key (right key).
  • Page 10: Chapter 5 Maintenance And Repair

    MD300I MD300I PULSE OXIMETER INSTRUCTION MANUAL Err 7: sensor emission diode or receipt diode damaged. CHAPTER 5 Maintenance and Repair Err 9: real time clock damaged. Warnings! Err 10: EEPROM chip damaged. • The advanced circuit inside the oximeter does not require periodic calibration and b) Possible problems and corresponding solutions maintenance, except replacing the batteries.
  • Page 11 MD300I MD300I PULSE OXIMETER INSTRUCTION MANUAL Our company will update the system software free of charge. • Not carefully read manual and so wrong operation lead to oximeter damage and failure. 5.4.2 Exempt and Limitation 5.4.5 User’s Special Request for Guarantee Time a) Our company isn’t responsible for such damage caused by force majeure.
  • Page 12: Appendix A Specifications

    MD300I MD300I PULSE OXIMETER INSTRUCTION MANUAL Data Cable Interface: DB9 (Connect to Pulse Oximeter); USB (Connect to PC) APPENDIX A Specifications Environment Requirements Notes: Operating Temperature: 5℃ ~ 40℃ Specifications may be changed without prior notice. Storage Temperature: -20℃ ~ +55℃...
  • Page 13: Appendix B Declaration

    Contact the carrier immediately if the shipping The model MD300I Pulse Oximeter is intended for use in the electromagnetic carton is damaged. specified below. The customer or the user of the model MD300I Pulse Oximeter should assure that it is such an environment. Emissions...
  • Page 14 MD300I Pulse Oximeter is used exceeds the applicable RF compliance level above, the below. The customer of the user of the MD300I Pulse Oximeter should assure that it is Pulse Oximeter should be observed to verify normal operation. If abnormal performance used in such an environment.
  • Page 15 Recommended separation distances between portable and mobile RF communications equipment and the MD300I Pulse Oximeter The model MD300I Pulse Oximeter is intended for use in an electromagnetic environment in which radiated RF disturbance are controlled. The customer or the user...