External, pH 4.0 (slope adjustment) & pH 7.0 (Cal. adjustment) Power Supply 006P DC 9V battery Power Consumption Approx. 2.0 mA Dimension 208 x 110 x 34 mm (L x W x H) Weight Approx. 380 g pH electrode pH plastic type, gel electrolyte with BNC connector AL10pH...
(0 mV at pH 7, based on 25°C working temperature). Ho- wever, because most pH electrodes do not exactly replicate the ideal electrode and because the working temperature may differ from 25°C, it is necessary to calibrate the instrument to ensure measurement accuracy of the combined pH electrode. AL10pH...
At the same time the display will show a Data Hold indicator. To release the display freeze function, press the Data. Hold Button again; the Data Hold indicator will disappear from the display. v) After taking a measurement, rinse the electrode in distilled water. AL10pH...
Electrodes are subjected to a natural ageing effect, which in- duces shift of the offset point and gradient of the electrode. For slowing down the ageing effect and for maintaining the performance and accuracy, please note the following considerations: AL10pH...
Page 7
- Water soluble contaminations: distilled water (tap water) - Chalk and hydroxide deposits: Hydrochloric acid (10 %) - Fats and oils: Warm water and mild detergents (dishwashing detergent) Attention: Do not use petroleum, ether or similar chemicals for plastic-covered electrodes. AL10pH...