3. FUNCTIONAL DESCRIPTION Figure 1 Calibration Adjust (HR, LR) Display Battery compartment/cover Power ON Button Stand Power OFF Button Electrode Handle Data Hold Button Conductivity Electrode 2 mS/cm Range Button Protective Cover 20 mS/cm Range Button Electrode Input Socket AL10Con...
Holding the conductivity electrode by its handle (Figure 1, 11), immerse it completely in the calibration solution. Shake the electrode to disperse air bubbles on it and allow measurements to stabilise. Adjust „Calibration Adjust LR“ (see above) until the display reads exactly 1.413 mS/cm. AL10Con...
Battery Compartment Cover (Figure 1, 9) on the rear of the meter. iii) Remove the battery, install a replacement one (006P DC 9V battery (heavy duty type), MN1604 (PP3) or equivalent) and replace the battery compartment cover. 7. ACCESSORIES 722250 Calibration solution 1413 μS/cm, 500 mL, traceable to N.I.S.T. AL10Con...