1.0 Introduction Information on the device versions Compared to the ATMOS® Cam 21, the ATMOS® Cam 31 has an additional image memory. The DV versions also have a USB-connection to conduct the video signal to the ATMOS® Cam 31 Cam 21 DV...
Ground wire connection Foot switch Alternating current On (feed-in, power connection) Observe operating instructions Off (feed-in, power connection) Follow operating instructions (blue) Control keys ATMOS® Cam 21 / 31 Endoscope Laryngoscope / Endoscope Usage: Examination of nose and Usage: Laryngeal examinations ears • Hysteroscopy •...
1.4 Function The ATMOS® Cam 21 / 31 consists of a camera head and the control unit. Both elements are connected detachably via a cable of 3 m length. The camera head includes a 1/3“ CCD colour image sensor with microlenses for connection to commercially available endoscopes with standard ocular objectives.
• Check the device and connection lines for defects prior to the use of the device. • Damaged cables must be replaced! Danger of explosion! • The ATMOS® Cam 21 / 31 is not designed for the use in medical areas with an explosion hazard or which are oxygenated. Explosion-hazardous areas may be caused by the use of flammable anaesthetics, skin cleansing products and skin disinfectants.
Page 8
Pay attention to the ambient conditions specified in chapter 9.0. The ATMOS® Cam 21 / 31 may not be operated with units not complying with the requirements of standard EN 60601-1 "Medical Electrical Equipment" and EN 60601-1-2 "Electromagnetic compatibility (Medical Electrical Equipment)".
EN-60601-1 For guaranteeing shortest connecting paths, the video components (recorder, printer, monitor, etc.) should be positioned near to the ATMOS® Cam 21 / 31. Please pay attention to operate all video components from one mains distributor. All additional equipment which is connected to the analogue and digital interfaces of the device must meet the requirements of relevant EN specifications (for inst.
Page 10
Forefront Buttons for selecting application part and light source Mains on/off switch Connection for camera head Connection for LED light source Rear side of the Cam 31 SD card Connection for HDMI video output Connection for video output (Y/C) Connection for video output (FBAS) Connection for rear side video input (Y/C)
Page 11
Rear side of the Cam 21 DV / 31 DV SD card Connection for HDMI video output Connection for video output (Y/C) Connection for video output (FBAS) Connection for rear side video input (Y/C) Connection for rear side video input (FBAS)
3.3 Digital monitor outlets The cameras ATMOS® Cam 31 DV and ATMOS® Cam 21 DV feature a digital monitor output with SXGA signal. The resolution of the monitor has to be at least 640 x 480 pixel. Please note that only the digital or analogue video outputs can be activated.
Page 13
Roadster) insertion like application part q.v. Chapter 4.3.1 Setting up and starting up...
If required, analogue peripheral video equipment (monitor, video printer, etc) must be adjusted to NTSC video norm. For further details please look at the corresponding operating instructions. 2. Optional 3. Valid for Cam 21 / 31 DV Y/C / S-VHS HDMI / DVI FBAS (composite) / VHS 5.
Page 15
3.6.1 Connection diagram ATMOS® Cam 21 / 31 Data digital analogue ATMOS® Strobo 21 LED (valid for cameras “Data“) FBAS-videosource e.g. ATMOS® Scope Computer S-video source (Y/C) • to Windows 7 e.g. ATMOS® i View (microscope) USB Streaming external microphone ATMOSoft REF 507.1806.0...
4.3 Camera head 4.3.1 Inserting / replacing the application part click 4.3.2 Focussing / image definition Pre-adjustment During the examination 4.3.3 Digital zoom Amplification and miniaturisation of examination image. Zoom steps: 1.0 x / 1.5 x / 2.0 x / 4.0 x. Operation...
4.4 Images-storage mode Only applicable for ATMOS® Cam 31 / DV / DATA. • Live mode: Displays live images of the examination. White balance and zoom are possible. • Memory mode: 1-image memory mode: One examination picture is saved on the SD card and displayed in full screen size.
4.5 SD card (Cam 31 / DV / Data) The SD card cannot be used if the ATMOS® Cam 31 / DV / Data is integrated in a treatment unit (e.g. ATMOS® S 61 Servant). If necessary remove the memory protection from the SD card and insert it with the printed side up.
4.7.1 White balance using the example of a halogen lamp The ATMOS® Cam 21 / 31 contains an automatic white balance, which allows you an uncomplicated and time-saving examination with best possible light configuration. However you can still perform a manual white balance. This manual adjustment will be maintained with this light source for subsequent applications.
5.0 Cleaning and care 5.1 General information on cleaning and disinfection • Do not spray the liquid directly on the camera. • If required, wipe the camera with a cloth moistened with a cleaning solution and disinfect the application parts which come into contact with the patient. For cleaning / disinfecting the optics please pay attention to the respective cleaning instructions.
Cleaning agents ..The application instrument may only be added to a completely dry camera head. 5.4 Chemical disinfection Avoid liquid residues in the snap-in connection of the video adaptor.. Upon drying avoid liquid residues and spots on the lens of the image sensor. All surfaces may be cleaned with the listed disinfectants.
6.0 Maintenance and Service Medical devices like the ATMOS® S 61 Servant vision need to be fail safe at any time. Therefore we recommend prior to every use: Further regular maintenance work is not required. Maintenance, repairs and period tests may only be carried out by persons who have the appropriate technical knowledge and are familiar with the product.
5. The device must be well padded and packed in suitable packaging. 6. Place the form QD 434 „Delivery complaint / return shipment“ and the respective decontamination certificate in an envelope. 7. Affix the envelope to the outside of the package. 8. Send the product to ATMOS or to your dealer. Maintenance and Service...
7.0 Troubleshooting Description Possible causes and measures No image • Are monitor and camera switched on? • Check monitor adjustment • Is the video cable correctly inserted? • Check cable connection from camera head to control device • Check fuses •...
Connecting cable (0.5 m) data transfer ATMOS Strobo 21 LED ® 507.5016.0 (included in scope of delivery ATMOS® Cam DATA) Audio cable (1.2 m) for transmission of the audio signal for video archiving 008.0858.0 (included in scope of delivery ATMOS® Cam DATA) Adapter plug for audio cable 008.0859.0...
Page 28
Laryngoscopes (ENT) Laryngoscope 70°, Ø 10 mm, working length 176 mm, 950.0209.0 autoclavable, without light guide 90°, Ø 10 mm, working length 174 mm, Laryngoscope autoclavable, 950.0210.0 without light guide 70°, Ø 8 mm, working length 166 mm, Laryngoscope autoclavable, without 950.0246.0 light guide...
Page 29
Flexible endoscopes (ENT) High-resolution Naso-Pharyngoscope Ø 3.8 mm, 950.0222.0 Working length 300 mm 0°, Angle of field of view 80° Depth of focus: 5 mm – infinite Angle 125° / 125° Supplied incl. aluminium carrying case, leakage tester ...
8.2 Options Endoscopy Trolley For integration of camera, stroboscope, endoscope management, monitor bracket, 950.0223.0 insulation transformer and foot control band. Endoscopy Trolley for Medical PC For integration of camera, stroboscope, endoscope management, monitor bracket, 950.0226.0 insulation transformer and foot control band. Insulation transformer 950.0225.0 Monitor bracket for endoscopy trolley...
10.0 Check-up / Disposal 10.1 Checking ATMOS devices The ATMOS devices are maintenance-free in the case they are used according to the operating instructions. At least every 24 months a repeat test of the electrical safety should be performed according to IEC 62353. ATMOS recommends an inspection according to the manufacturer‘s specifications.
Halogen light source Flexible hysteroscope Xenon light source ATMOS® i View Camera control Only applicable for AT- unit MOS® Cam 21 / 31 DV Video FBAS HDMI Video processing and archiving Video monitor in standard video format (S-VHS, FBAS) ...
Guidance and manufacturer's declaration - electromagnetic emissions The ATMOS® Cam 21 / 31 is intended for use in the electromagnetic environment specified below. The customer or user of the ATMOS® Cam 21 / 31 should ensure that it is used in such an environment. Electromagnetic Environment - Emissions Test...
Page 36
Guidance and manufacturer's declaration - electromagnetic immunity The ATMOS® Cam 21 / 31 is intended for use in the electromagnetic environment specified below. The customer or user of the ATMOS® Cam 21 / 31 should ensure that it is used in such an environment. IEC 60601- Test Electromagnetic Environment - Immunity Test Compliance Level...
Page 37
If the measured field strength at the location where the ATMOS® Cam 21 / 31 is used exceeds the above compliance level, the ATMOS® Cam 21 / 31 is to be observed to verify the intended use. If abnormal performance characteristics are noted, additional measures might be necessary, e.
Page 38
Recommended separations between portable and mobile RF Communications equipment and the ATMOS® Cam 21 / 31 The ATMOS® Cam 21 / 31 is intended for use in electromagnetic environment in which radiated disturbances are controlled. The customer or user of the ATMOS® Cam 21 / 31 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF Communications equipment and the ATMOS®...