Conductivity; Measuring; Measuring The Conductivity - wtw Multi 9420 Operating Manual

Digital meter for ids sensors
Table of Contents

Advertisement

Conductivity

Selecting the
displayed
measured parameter
Stability control
(AutoRead)
& HOLD function
78
8
Conductivity
8.1

Measuring

8.1.1 Measuring the conductivity

The sensor connection and the USB-B (device) interface are gal-
vanically isolated. This facilitates interference-free measurements
also in the following cases:
 Measurement in grounded test samples
 Measurement with several sensors connected to one Multi 9420
in one test sample
1.
Connect the conductivity sensor to the meter.
The conductivity measuring window is displayed.
The measuring cell and cell constant for the connected IDS conductivity
sensor are automatically taken over.
If necessary, press <M> to select the measured parameter ϰ.
2.
3.
Immerse the conductivity sensor in the test sample.
Info
03.04.2013 08:00
You can switch between the following displays with <M>:
 Conductivity [μS/cm] / [mS/cm]
 Resistivity [Ω·cm] / [kΩ·cm] / [MΩ·cm]
 Salinity SaL [ ]
 Total dissolved solids TDS [mg/l] / [g/l]
The factor to calculate the total dissolved solids is set to 1.00 in the factory. You
can adjust this factor to meet your requirements in the range 0.40 ... 1.00. The
factor is set in the menu for the parameter TDS.
The stability control function (AutoRead) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproduc-
ibility of measured values.
The measured parameter flashes on the display
 as soon as the measured value is outside the stability range
 when the automatic Stability control is switched off.
Multi 9420
ba75964e04
04/2015