40021
0020
40023
0022
40025
0024
40027
0026
Two bytes read and write(optional) address(function code:03,06)
40101
0100
40102
0101
40103
0102
40104
0103
40105
0104
40106
0105
40107
0106
40108
0107
40109
0108
40110
0109
40111
0110
40112
0111
40113
0112
GM8802-F2110101
Reserved bit
Reserved bit
State from channel no. 1~2
GM8802-F2 code
Channel no.1 digital filter
Channel no. 1 stable range
Channel no. 1 stable time
Channel no. 1
zero-tracking range
Channel no. 1
zero-tracking range
Channel no. 1 zeroing
range
Channel no. 1 unit
Channel no. 1 deciaml
digits
Channel no. 1 minimum
division
Channel no. 1
anti-vibration parameters
Channel no.2 digital filter
Channel no. 2 stable range
Channel no. 2 stable time
GM8802F-2 Field Mounting Transmitter
Fixed data:0x7F+'R'+'E'+'V'
Channel no. 1 sign:
bit0:0 unstable / 1 stable
bit1:0 normal / 1 overflow
bit2:0 nonezero / 1 zero
bit3:0 positive sign / 1 minus sign
bit4:0 AD normal / 1 AD error
bit5:0AD shut / 1AD start
Channel no. 2 sign:
bit6:0 unstable / 1 stable
bit7:0 normal / 1 overflow
bit8:0 nonezero / 1 zero
bit9:0 positive sign / 1 minus sign
bit10:0 AD normal / 1 AD error
bit11:0AD shut / 1AD start
bit12~bit31:0(reserve)
GM8802-F2 read code:
0x30324632(ASCII code is 02F2)
Please see Character 4 for details.