Parameter Table; Operation Parameters - General Measure GM8802F-2 User Manual

Field mounting transmitter
Table of Contents

Advertisement

4 Parameter table

Parameters in every channel are same, but also can be set individually.
4.1 Calibrate parameters
Parameters
Unit
Decimal point
Minimum division
Maximum capacity

4.2 Operation parameters

Parameters
Filter grade
Stability range
Stability time
Zero-tracking
range
Zero-tracking
time
Zeroing range
Anti-vibration
parameters
GM8802-F2110101
Initial values
10000
Initial
Range
values
4
0-9
2d
1-9d
1.0S
0.1-1.0S
0d
0-9d
0.5,1.0,1.5,
1.0S
2.0
20%
1%-99%
4
0-99
GM8802F-2 Field Mounting Transmitter
1
0
2
1,2,5,10,20,50
≤Mini division×50000
0:none
9:strongest
If present weight change at
stability range in stability time, then
transmitter confirm the weight
value is stable.
If present weight is at zero-tracking
range in zero-tracking time, the
transmitter track zero, but not track
zero when Zero-tracking range is
0.
Percent of maximum capacity.
0 means no anti-vibration function
Range
0,1,2,3
0,1,2,3,4
Description
9

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents