Chapter 6: Specifications - Mantracourt ALA5 User Manual

2-wire 4-20ma atex load cell amplifier
Table of Contents

Advertisement

Chapter 6: Specifications

Table 6.1 ICA5ATEX – 4-20mA 2 wire Specifications
Environmental
Parameter
Supply voltage Range
Operating Temperature
Range
Storage Temperature
Range
Reverse polarity
Protection
Measurement
Parameter
Bridge Excitation
Bridge Impedance
Bridge Sensitivity
Output load
Bandwidth
'Zero' adjustment
'Span' adjustment
Linearity
Temperature stability
'Zero' Temperature
Stability
'Span' Temperature
Stability
FR=Full Range (16mA)
Note 1: 1000 Ohm load cell – Typically 0.53V for 350 Ohm cell
Note 2: Set by calibration resistor
Note 3: Limited by the 300Ω barrier impedance
Note 4: 1000 Ohms load cell – change R1 to suit other load cell impedances.
Note: Recommended bridge impedance is 1,000 Ohms
Note: The voltage between either of the power supply connections and the load cell shield should not exceed 50V.
Any leakage will be greater than 10M Ohms.
Table 6.2 ILEATEX Enclosure Specifications
Parameter
Cover material
Body material
Threaded entries
Cable entries
Environmental rating
9
Mantracourt Electronics Limited ALA5, ILEAtex and ICA5Atex User Manual
Minimum
Typical
9.0
24
-40
-
-40
-
-30
-
Minimum
Typical
1.05
1.11
350
1000
0.5
2.5
-
-
DC
-
-
±2
-
±8
-
0.02
-
0.001
-
0.007
Minimum
Stainless steel type 304
Stainless steel type 316 or 17-4
PG7 / 20 TPI
PG7 EMC glands
(e.g. Jacob 50.007/EMV 14mm AF)
IP67
Maximum
Units
28
Volts
85
Deg C
85
Deg C
-
Volts
Maximum
Units
1.16
Volts
5000
Ohms
55
mV/V
300
Ohms
1000
Hz
-
%FR
-
%FR
-
%FR
0.005
+/-%FR/Deg C
0.014
+/-%FR/Deg C
Notes
Notes
Note 1
Note 2
Note 3
Note 4
At 2.5mV/V
At 2.5mV/V

Advertisement

Table of Contents
loading

Table of Contents