3. Gently ease the base out of the cover
ensuring that the cable slides through
the loosened gland to avoid any strain on
the internal connections.
Calibration:
1. Apply the low calibration conditions (weight, force or mV/V). Set the output to 4mA using the Zero
potentiometer.
2. Apply the known high calibration conditions (ideally between 75% and full scale) and adjust the Span
potentiometer to give the required output current for the known input.
i.e. 16mA for 4-20mA final calibration with 75% input or 20mA if 100% input.
When re-assembling the ALA5 it is imperative that the 'O' is fitted correctly and the cable gland re-tightened to
maintain a gas-tight seal.

Chapter 4: Changing the Gain and Offset Ranges

The ICA5ATEX amplifier is supplied un-calibrated but optimized for a sensitivity of 2.5mV/V with a 1k bridge.
To accommodate other sensitivities the gain resistor, 'R2' (see below) can be changed to a value derived from the
following formula.
Figure 4.1
N.B. a high quality resistor (e.g. 1% 25-50 ppm metal film) should be used for best performance. It may be necessary
to use an E96 value for optimum trim range: -
Mantracourt Electronics Limited ALA5, ILEAtex and ICA5Atex User Manual
4. The ICA5ATEX amplifier will now be
exposed allowing access to the two
trimming potentiometers.