1
Before Use
Thank you for purchasing a
Labcold
IntelliCold® Pharmacy Refrigerator. Your fridge has been
purposely designed and precision built for the safe storage of vaccines and other temperature
sensitive pharmaceuticals.
Although great care has been taken in the manufacture of your refrigerator, damage can occur
in transit. Before using for the fi rst time please remove all packaging and give your fridge a
thorough visual inspection. If any damage is suspected, please contact your retailer immediately
upon receipt of the product. Failure to do so may invalidate your warranty.
Setting Up
Place your fridge in the location in which it is to be used.
Do not place your fridge near a direct heat source or direct sunlight and leave at least 6cm all
round the unit for adequate ventilation. Models designed to be wall mounted do not need extra
space at the back because the cooling system is in the sides. Failure to do so may eff ect the
performance of your refrigerator.
Clean your fridge with mild soapy water to remove any dust that may have accumulated during
manufacture and dry thoroughly.
Ensure your fridge is level by adjusting the feet at the bottom of your fridge.
For models RLDF/G0501A, 1010A and 1510A, insert shelves by placing clips in the racking and
resting the shelf on top. When positioning the shelves, please remember to keep adequate room
for air circulation. Failure to do so may cause hot spots in the fridge.
YOUR REFRIGERATOR SHOULD NOW BE LEFT FOR AT LEAST TWO HOURS BEFORE
CONNECTING TO POWER FOR THE FIRST TIME
A Closer Look
Front Panel Display
Door lock
Led Light
Fan
Remote Alarm
and Accessory
Sockets
(back of fridge)
Adjustable Feet
Alarm Battery Back Up
Insert 4 AA long life alkaline batteries (not supplied) in the holder
located on the right side of the unit
Batteries will only power the controller not the fridge.
Remote Alarm and
Remote alarm and
Labcold
Labcold accessories
back of the fridge. If you have a
sockets
can be connected to the 2 pin connector on the left. The 4 pin
connector socket is for the
The maximum rating for the connection is 1A 30Vdc.
DO NOT CONNECT TO MAINS VOLTAGE
2
Battery Holder
(side of the fridge)
SD Card
slot
Serial Plate
Adjustable Shelves
Maximum Load
10kg
accessory sockets are located at the
Labcold
external alarm kit this
Labcold
accessory kit.
Need help?
Do you have a question about the IntelliCold and is the answer not in the manual?
Questions and answers
Fridge RLDF 1510A dEF display on display
@Khurshid is this an error code?
What does Y24 code mean
My fridge is not working, there is what I can only assume to be an error code on the front of the fridge that has the letter t in the second column and nothing else showing however this error code does not seem to show up on any of my searches, is my fridge totally broken or can this be fixed?
I have a 5 flashing on the display panel on the Left hand side of the screen. I pressed set and it has disappeared . What does this mean please?
what does code D14 mean on fridge
d-0 what meaning
@Shameer door open
HI i **** getting a messsage "b6tt" or "batt" I cant seem to find what this means
The date on the digital display is showing an old date (somewhere in 2024) and we are afraid if we set the date correctly, the data in the controller will be overwritten. And now the displaying is bleeping and show 'PoFF' and we want to download the data. How to fix this problem and reserve the data already recorded in the controller?
FRIDGE BEEPING AND SAYING t2h
The fridge broke coldchain after restocking at 14:15 hrs on the Monday and was found to be current temp was 6.1, minimum temp was 3.7 and maximum was 10.4 at 18:14pm but was not rest until Tuesday morning at 0800hrs when it was within normal range except for maximum still at 10.4; the fridge was then reset within cold range. The fridge was never opened before it was reset. Can you let me know if it is still safe to use the vaccines.