2000KL ASSEMBLY INSTRUCTIONS (CONTINUED)
Assembling the 2000KL Scale (Refer to Figures 1 - 12)
Caution: It is recommended that the assembly be performed by two persons.
Assembling the Patient Lifter
1. Remove the lifter base from the carton and place the lifter base on a flat, level and dry surface
(see Figure 2).
2. Place the stretcher holder onto the lifter base (see Figure 3).
3. Remove black plastic protective caps from lifter mast (see Figure 3).
4. Insert the bottom of the lifter mast and hydraulic pump assembly into the mast sleeve (through
the stretcher holder) on the lifter base (see Figures 2 & 3).
5. The notch on the end of the lifter mast fits over the round tube at the bottom of the mast
sleeve (see Figure 3).
6. Position and lock the lifter mast into the mast sleeve. When assembled correctly, the lifter
mast will not be able to rotate (see Figure 3).
7. Check that the lifter mast is locked into position, and that the hole in the lifter mast is aligned
with the threaded hole in the mast sleeve (see Figure 3).
8. Screw the knob into the threaded hole in the mast sleeve. Tighten to lock the lifter mast to the
lifter base (see Figure 3).
Keep lifter mast and lifter base locked at all times, except when removing lifter mast from
Note:
the lifter base for storage.
9. Locate the NC 3/8 – 20 x 2 ½" hexagon head screw (see Figure 1, Fastener "I").
10. Insert the hexagon head screw through the stretcher retaining strap eyelet hole. The screw
can be inserted through either side of the strap (see Figure 4).
11. Locate the self lock nut (see Figure 1, Fastener "V").
12. Using the same hexagon head screw, assemble the lifter boom and the display support to the
lifter mast. Thread the screw through all three parts, and secure the screw with the self lock
nut (see Figure 4).
Note: Ensure that the
13. Locate the NC 3/8 – 20 x 1 ¾" hexagon head screw and self lock nut (see Figure 1, Fastener
"II" and "V").
14. Insert the end of the hydraulic pump shaft into the lifter boom's bracket, and align the
assembly holes (see Figure 4).
15. Thread the hexagon head screw through the lifter boom bracket and the hydraulic pump shaft
hole. Secure the screw with the self lock nut (see Figure 4).
16. Insert the base adjustment lever into its socket at the back of the lifter base. Check to see that
the positioning ball on the outside of the base of the adjustment lever fits securely in the
socket (see Figure 5).
lifter boom hooks are facing down during patient lifter assembly.
8