CAUTION AND WARNING

To prevent injury and damage to your scale, please follow these instructions very
carefully.
Assemble and operate the scale per the enclosed user instructions.
Operate this scale exclusively with the AC adapter provided by Health o
meter. Use of an unspecified adapter will void the warranty, and can pose
a serious safety hazard.
Prior to using this scale, inspect the AC adapter cord for cracking /
fraying, or for broken / bent plug prongs.
Prior to using this scale, ensure that the AC adapter is plugged into an
outlet with the rated voltage appropriate for operation.
Ensure that the AC adapter outlet is wired to a circuit breaker or other
protected power source.
Unplug the AC adapter, and carefully store both the adapter and adapter
cord before moving the scale.
Ensure that the AC adapter and scale do not come in contact with liquids,
excessive temperature, or excessive humidity.
Operating this scale at voltages and frequencies other than specified can
damage the equipment and will void the warranty.
Do not exceed recommended weight limit of 400 lb / 181 kg for this scale.
For accurate weighing, this scale must be placed on a flat, stable surface.
For accurate weighing, the casters on the lifter base must be placed on a flat and
stable surface with the weighing mechanism not touching the floor.
For accurate weighing, verify before each use the proper operation
according to the procedure described in this manual.
Do not use in the presence of flammable or explosive materials.
For accurate weighing, if the "Lo bAtt" indicator activates, replace the
batteries or switch to the AC adapter as soon as possible.
If the scale becomes damaged, it should not be operated until properly serviced. All
repairs should only be performed by authorized Health o meter
service personnel.
This scale is designed for static weighing of patients only. No scale should be used
for patient transfer. Pelstar, LLC assumes no responsibility for patient injury or
scale damage should this caution be ignored by the scale operator.