Model 2000KL ® Thank you for your purchase of this Health o meter Professional product. Please read this manual carefully, and keep it for easy reference or training. TABLE OF CONTENTS Caution and Warning....................3 Specifications for This Scale ..................5 Certifications / Connectivity / Disposal ..............6 Assembly Instructions 2000KL Scale ...............7...
AC adapter as soon as possible. If the scale becomes damaged, it should not be operated until properly serviced. All ® repairs should only be performed by authorized Health o meter Professional service personnel. ...
Page 4
CAUTION AND WARNING Do not leave patients unattended while patient is being weighed on the scale. When replacing / inserting batteries, be sure to use all new batteries. For maximum safety, ensure stretcher is positioned above a bed or a similar surface.
SPECIFICATIONS FOR THIS SCALE General The Model 2000KL stretcher scale uses highly sophisticated microprocessor technology. Each precision scale is designed to provide accurate, reliable and repeatable weight measurements. In addition, each scale is designed to provide the user with features that make the weighing process simple, fast, and convenient.
This scale is EMR-EHR compatible. For more information regarding how to connect this scale to ® other electronic devices, please contact Health o meter Professional Scales Technical Support, available during normal business hours at 1-800-638-3722.
Page 7
2000KL ASSEMBLY INSTRUCTIONS Before Assembly Each 2000KL stretcher scale is shipped disassembled in (3) separate cartons. Carefully inspect all cartons for shipping damage before unpacking. If damage is found, contact your shipper or a ® Health o meter Professional representative immediately at 1-800-815-6615. Claims must be filed with the shipper as soon as possible after receipt of the packages.
Page 8
2000KL ASSEMBLY INSTRUCTIONS (CONTINUED) Assembling the 2000KL Scale (Refer to Figures 1 - 12) Caution: It is recommended that the assembly be performed by two persons. Assembling the Patient Lifter 1. Remove the lifter base from the carton and place the lifter base on a flat, level and dry surface (see Figure 2).
Page 9
2000KL ASSEMBLY INSTRUCTIONS (CONTINUED) 17. To adjust the lifter base with casters, unlock the legs by pulling back on the adjustment lever. Move the lever to the right in order to widen the base (see Figures 5 & 8). 18. A caster lock is provided on the lifter base to prevent movement. To lock caster, step on caster lock lever on side of caster.
Page 10
2000KL ASSEMBLY INSTRUCTIONS (CONTINUED) 29. Use the hexagon shoulder screws (III) to connect the two stretcher hoops to the load cell beam assembly and securely tighten. Make sure that the stretcher hooks face each other in the inward position, toward the stretcher canvas (see Figures 1 & 11).
2. Remove the battery cover on the side of the display assembly by unscrewing the (2) battery cover screws (see Figure 13). 3. Replace or install the batteries with new units. Health o meter recommends using an alkaline or lithium battery, rated for 130°F (54°C) operation. Do not use Zinc-Carbon batteries.
QUICK START / OPERATING INSTRUCTIONS Keypad Figure 14 Function Description Turns scale on and off. ON/OFF Zeros the scale prior to a weighing. ZERO Holds the weight value of the patient on the display until the button is HOLD/RELEASE pressed again to clear the value. Also moves the cursor from left to right ►...
Page 20
The Tare Function automatically performs this subtraction. The 2000KL scales allow tare weight to be entered manually by using the keypad, or automatically, as outlined in this section.
Page 21
QUICK START / OPERATING INSTRUCTIONS (CONT) 8. The weight of the patient will appear on the scale. 9. With no weight on the scale, the tare value will remain on the display until changed, cleared, or until the scale is turned off. 10.
Page 22
QUICK START / OPERATING INSTRUCTIONS (CONT) 5. Once a weight has been displayed, while the patient is still on the scale, Health o meter recommends performing a second weighing by pressing the REWEIGH button. 6. The weight may be held on the display by pressing the HOLD/RELEASE button. To release, press the HOLD/RELEASE button again.
Page 23
QUICK START / OPERATING INSTRUCTIONS (CONT) Note: When weighing in pounds, the height is displayed in inches. When weighing in kilograms, the height is displayed in centimeter. 8. Once height is keyed in, press the ENTER button. 9. The display will show the patient’s BMI. 10.
Maintenance The following pages provide instructions for maintenance, cleaning, calibration, and troubleshooting for the 2000KL scale. Maintenance operations other than those described in this manual should be performed by qualified service personnel. Caution: Before first use, or after long periods of non-use, check the scale for proper operation and function.
This scale has been factory calibrated, and does not require calibration prior to use. For maintenance, the scale can be calibrated. Calibration of your 2000KL scale is performed using kilograms (kg) or pounds (lb), according to the units of measure used upon entering into the calibration path.
TROUBLESHOOTING Troubleshooting Before contacting service personnel, refer to the following instructions to check and to correct any failures. Symptom Possible Cause Corrective Action Scale does not turn on 1. Dead battery 1. Replace batteries 2. Faulty electrical outlet 2. Use a different outlet 3.