Operation of specimen holder
1.4.3
LEO 912 OMEGA

Lock-in specimen rod

SPECIAL NOTE:
The normal position of the specimen rod is in the high vacuum.
Take it only out for specimen exchange.
Slide specimen rod carefully into goniometer as far as it will go.
Turn specimen rod carefully clockwise through 90° all the way to
prevacuum position.
– Airlock valve opens (click).
– Red LED STOP is on. Specimen rod pre-pumped in approx. 20 sec.
LED STOP goes out.
Wait until LED STOP is out.
Pull specimen rod back to stop and turn it carefully through further 90°;
push rod slowly into the microscopy position
Insertion of specimen rod
Preevacuation concluded
Microscope Operation
Preevacuation
Microscopy