Centration of spectrometer slit aperture

1.9
1.10
LEO 912 OMEGA
Centration of spectrometer slit aperture
Switch to SPEC with button IMAGE/SPEC in the field SPECTROMETER.
– Spectrum caustic close to the center of the screen.
Center tip of the caustic to the index point with encoder in the field
SPECTROMETER.
Introduce slit aperture with the aperture drive and adjust the slit width
that the spectrum caustic is within the slit.
– The tip of the caustic should not be too close to the corresponding side of the
slit.
SPECIAL NOTE:
If elastic brightfield is used only, the slit width might be increased so that the caustic
tip is further off the slit edge.
Switch to IMAGE mode with button IMAGE/SPEC.

Correction of imaging astigmatism

Select a magnification between 125.000x and 250.000x with buttons
MAGNIFICATION ∆ ∇.
Adjust brightness by changing the emission current with encoder
CURRENT or the illumination aperture angle with the buttons
ILL APERTURE ∆ ∇.
Note, the illumination angle should not be larger than 1.6 mrad. Larger
illumination angles decrease the image quality.
Through-focus specimen with encoder OBJECTIVE (Grain of a carbon film
or embedding resin).
– The image is astigmatic, if the grain of the carbon film has a privileged direction
in underfocus and is turned by 90° in overfocus.
Switch to STIG with button SHIFT/STIG in the field IMAGE.
Compensate astigmatism with the encoders X and Y in the field IMAGE.
– The astigmatism is compensated, if there is no privileged direction of the grain
in under- and overfocus.
Switch back to SHIFT with button SHIFT/STIG in the field IMAGE.
Adjust brightness and magnification to the desired values.