Microscope Operation
1.7
1.8
8
Activate function ADJUST AIS LOWER and center AIS aperture with
encoders X, Y in the field ILLUMINATION.
Switch to DIFF mode and focus the diffraction spot with encoder
OBJECTIVE
in the field FOCUS.
Activate function ADJUST STIG CORR and correct astigmatism with
encoders X, Y in the field ILLUMINATION.
– The diffraction spot may move by the astigmatism correction. If the spot is
outside of the objective aperture, continue with the next step first and then
repeat this one.
Activate function ADJUST TILT CORR and shift diffraction spot to the
index
point (Center of the objective lens aperture) with encoders X and Y in the
field ILLUMINATION.
Switch to IMAGE mode.
Adjust the AIS system at 31.500x, 16.000x and 8.000x in the same way.
Check the complete adjustment and optimize it.

Centration of objective lens aperture

Select diffraction mode with button IMAGE/DIFF in the field IMAGE.
Adjust the brightness of the diffraction image with the buttons
ILL APERTURE ∆ ∇.
Select a camera length of CL = 580 mm with the buttons
MAGNIFICATION ∆ ∇.
Center the diffraction spot to the index point with X and Y in the field
IMAGE.
Focus the diffraction image with encoder OBJECTIVE in the field FOCUS.
Insert the desired objective aperture and center it to the diffraction spot.
Switch back to IMAGE mode with button IMAGE/DIFF in the field IMAGE.

Centration of spectrometer aperture

Insert entrance aperture (1st clickstop) and center it with the aperture
drives to the large screen.
Centration of objective lens aperture
LEO 912 OMEGA