Health O Meter 524KL User Instructions

Health O Meter 524KL User Instructions

Digital pediatric seat scale
Hide thumbs Also See for 524KL:

Advertisement

Model 524KL / 524KG
Digital Pediatric Seat Scale
User Instructions
th
PELSTAR, LLC
9500 West 55
St. McCook, IL 60525-7110 USA
www.homscales.com
© Pelstar, LLC 2018
P/N UM524KL_524KG Rev20180227
1

Advertisement

Table of Contents
loading

Summary of Contents for Health O Meter 524KL

  • Page 1 Model 524KL / 524KG Digital Pediatric Seat Scale User Instructions PELSTAR, LLC 9500 West 55 St. McCook, IL 60525-7110 USA www.homscales.com © Pelstar, LLC 2018 P/N UM524KL_524KG Rev20180227...
  • Page 2: Table Of Contents

    Model 524KL / 524KG ® Thank you for purchasing this Health o meter Professional product. Please read this manual carefully, and keep it for easy reference or training. TABLE OF CONTENTS Cautions and Warnings ..................... 3 Specifications ......................4 Certifications / Connectivity / Disposal ..............5 Assembly Instructions ....................
  • Page 3: Cautions And Warnings

    ® The Health o meter Professional 524KL / 524KG scale is intended to be used in a professional medical environment by trained medical staff. This product was designed to weigh infants that are sitting in the weighing seat. Do not modify the product or use it for anything other than its intended purpose.
  • Page 4: Specifications

    Scale Specifications Capacity and Resolution 524KL: 50 lb / 23 kg; 0 - 20 lb / 0.2 oz; 20 - 50 lb / 0.5 oz; 0 - 9 kg / 5g; 9 - 23 kg / 10 g 524KG: 23 kg; 0 - 9 kg / 5g; 9 - 23 kg / 10 g...
  • Page 5: Certifications / Connectivity / Disposal

    For reliable transmission of weight data, this scale is designed to connect to a computer, ® monitor, or other electronic data device via Health o meter Professional Connectivity Solutions. This connectivity helps minimize transcription errors when recording patient data. For further information regarding how to connect this scale to other electronic devices, please contact ®...
  • Page 6: Assembly Instructions

    ASSEMBLY INSTRUCTIONS Before Assembly Each 524KL / 524KG Digital Pediatric Seat scale is shipped disassembled in one carton. Carefully inspect the carton for shipping damage before unpacking. If damage ® is found, contact your shipper or a Health o meter Professional representative immediately at 1-800-815-6615.
  • Page 7 ASSEMBLY INSTRUCTIONS (CONTINUED) 1. Carefully remove the contents of the Screws and washers carton and place each part on a flat, level, and dry surface. Remove the 4 column mounting screws and washers from the bottom of column and set aside for assembly step 5. 2.
  • Page 8 ASSEMBLY INSTRUCTIONS (CONTINUED) 6. Locate the two black seat mounting brackets in the carton. Unscrew and remove the black knobs on the mounting brackets. Note: The brackets are marked as bracket C and bracket D. The tabs on bracket C are shorter than bracket D. 7.
  • Page 9: Powering The Scale

    Locking tab Display assembly Battery cover ® 2. Replace or install 6 new AA batteries. Health o meter Professional recommends using an alkaline or lithium battery, rated for 130°F (54°C) operation. Do not use zinc-carbon batteries. 3. Re-attach the battery cover to the display assembly.
  • Page 10: Set Up Instructions

    SET UP INSTRUCTIONS AUTO-OFF FUNCTION – MODELS 524KL AND 524KG As an energy saving feature, the scale will automatically turn off after 2 minutes of inactivity. By default, this function is enabled. Note: The Auto-off function is only active when the scale is being powered by batteries.
  • Page 11: Operating Instructions

    OPERATING INSTRUCTIONS WEIGHING A PATIENT – MODELS 524KL AND 524KG 1. Make sure there is nothing on the weighing seat. 2. Press the ON/OFF button to turn the display assembly on. “88:8:8.8” will appear on the display followed by zeroes.
  • Page 12: Maintenance

    4. Do not use abrasive cleaners. Replacement Parts ® Parts may be available if a piece of the scale needs to be replaced. Contact Health o meter Professional Scales Customer Service at 1-800-815-6615 to inquire about the availability of these replacement parts.
  • Page 13: Calibration

    DISPLAY SHOWS “CAL” followed by With the scale off, press and hold the LB/KG button (524KL) or the “XXXX” blank blue button (524KG) and ZERO buttons at the same time, then press the ON/OFF button to turn on the scale. “CAL” will briefly appear on the display, followed by four numbers indicating that you can begin calibration.
  • Page 14: Troubleshooting

    TROUBLESHOOTING Troubleshooting Before contacting service personnel, refer to the following instructions to check and to correct any failures. Symptom Possible Cause Corrective Action Scale does not turn on 1. Dead battery 1. Replace batteries 2. Faulty electrical outlet 2. Use a different outlet 3.
  • Page 15: Warranty

    ® Health o meter Professional products are manufactured, designed, and owned by Pelstar, LLC. ® We reserve the right to improve, enhance, or modify Health o meter Professional product features or specifications without notice. © Pelstar, LLC 2018 P/N UM524KL_524KG Rev20180306...

This manual is also suitable for:

524kg

Table of Contents